18.2
TEMPERATURE ALARMS FROM THE COOLER
Error message
Lowtemp.: 1
Hightemp: 1
T> Max: 1
T <min:1
Lowtemp.: 2
Hightemp.: 2
T> Max.: 2
T <Min: 2
The following screen appears for channel 2 if the control temperature (+ 10 °C alarm hysteresis) is exceeded:
This prevents the heated line or the heated probe from overheating due to a defect! The triggered alarm (red LED)
must be acknowledged by pressing any key. If the error is not corrected, the error screen comes back after 60 s.
18.3

FLOW DISTURBANCE AND LIQUID ALARMS

Error message
FA-Warn: 1, 2
LA-Alarm: 1, 2
LA-Break: 1, 2
In case of liquid alarm and activated LA-Reset (see chapter 17.6) the following screen occurs:
34
Alarm
Warning
x
-
x
-
x
x
-
x
x
-
x
x
Alarm
Warning
-
x
x
-
x
-
CSS-VC-1.6 | 1.02.01
LED
Error
red
Low temperature cooler
red
High temperature cooler
red
Temperature sensor range exceeded =>
sensor defective
red
Temperature sensor measuring range fallen
below => sensor defective
yellow
Low temperature channel 2
red
High temperature channel 2
red
Temperature sensor range exceeded =>
sensor defective
red
Temperature sensor measuring range fallen
below => sensor defective
LED
Error
yellow
Flow disturbance channels 1 and/or 2
red
Liquid alarm channel 1and/or 2
red
Cable break liquid sensor channel 1 and/or 2