M&C CSS-VC1 Instruction Manual page 52

19", wall mount or portable in carrying case with software version
Hide thumbs Also See for CSS-VC1:
Table of Contents

Advertisement

Cancellation of alarms
Event message
T1 OK
T2 OK
LA1 OK
LA2 OK
T1-Temp <max
T2-Temp <max
T1-Temp> min
T2-Temp> min
Cancellation of warnings
Event message
T2 L warning OK
Device OK
FA1 OK
FA2 OK
Pump1 OK
Pump2 OK
Calib. OK
Service OK
52
Significance
T1 is in the control range or Pt100 is OK
T2 is in the control range or Pt100 is OK
Humidity sensor 1 is OK or has no humidity
Humidity sensor 2 is OK or has no humidity
T1 <99.9 °C
T2 <248 °C
T1-Temp> min
T1> -70 °C
T2-Temp> min
T2> -50 °C
Significance
T2 now has no low temperature
Device within the operating temperature (1 °C> T <50 °C)
Flow channel 1 OK
Flow channel 2 OK
Pump1 running after alarm cancelled
Pump2 running after alarm cancelled
Cancelling of an internal error after returning the device
Service reset carried out
CSS-VC-1.6 | 1.02.01