M&C CSS-VC1 Instruction Manual page 32

19", wall mount or portable in carrying case with software version
Hide thumbs Also See for CSS-VC1:
Table of Contents

Advertisement

To discard the changed settings, keep the ESC key
After saving or discarding the settings, you will automatically return to the start screen.
If the unit is disconnected from the mains voltage while it is in user setup, all changes made are also rejected.
The following values can be set:
Service interval
Language
Ctrl. temp.
HSL
LA-Reset
32
pressed for 5 seconds.
=
(0 (-----) - 365 days/default setting 180 days/off). After the set time has
elapsed, the warning LED (yellow) LIGHTS UP and the following
appears in plain text: "Service required". In addition, the pending
warning is also reported via the individual alarm connection (see
Section 12.4.2).
=
(German, English, French)
=
(0 - 7 °C [32 °F – 44.6 °F] (/default setting 5 °C [41 °F])
=
(0 – 240 °C [32 – 464 °F], default setting 180 °C [356 °F])
=
(x = reset necessary, ----- = no reset necessary)
CSS-VC-1.6 | 1.02.01
www.mc-techgroup.com