Troubleshooting; Faults (With Messages) - Zeiss VISULAS green Manual

Hide thumbs Also See for VISULAS green:
Table of Contents

Advertisement

Documentation set
VISULAS green
Error message
Door switch
Laser stop switch
Connection error of physician's
safety filter
Fiber not connected
Trigger not connected
Temperature monitoring
Power warning
Trigger pressed.
000000-2096-101-DokS-en-GB-150322

10 Troubleshooting

10.1 Faults (with messages)

Faults are indicated as system information on the control unit display.
The user is also alerted by short audible signals.
Note: Follow the instructions given.
Warning message: Remedy the fault using the troubleshooting tables
below.
If a fault occurs, which cannot be remedied by using the following
u
tables, please contact ZEISS Service, citing the error number
displayed.
Label the system as being out of order, remove the key from the
u
key switch and disconnect the power plug.
Cause
The door contact is open.
The laser stop switch is locked.
The connection to the slit lamp
or the physician's safety filter
was interrupted.
No optical fiber detected.
No trigger for laser detected.
Temperature within the laser
console is too high.
The internal monitoring
systems have identified that
the actual laser power output
deviates from the target power
output selected on the control
panel by more than 20%.
For safety reasons, it is not
permitted to switch to
treatment mode while the
trigger is pressed.
10.1 Faults (with messages)
Remedy
Shut the door or check that the
u
shorting plug is correctly plugged into
the yellow socket.
Unlock the laser stop switch.
u
Check that the slit lamp and the OPMI
u
(ENDO physician's safety filter) are
correctly connected and properly
inserted.
Check that the optical fiber is properly
u
inserted and screwed in tightly.
Check that the foot switch and the
u
slit lamp are correctly connected and
properly inserted.
Make sure that the vent slots on the
u
laser console are not covered.
Ensure that the ambient temperature
u
is below 35 °C.
Contact ZEISS Service if the message
u
occurs repeatedly.
To continue, release the trigger.
u
10 Troubleshooting
109 / 136