2.4 Measures and duties of the operator
Electrical safety
Switch the system off every time before disconnecting it from the
u
power supply, or if you are not going to use the system for any
length of time. Also disconnect the system from the power supply
before cleaning surfaces or accessories.
Insert the power plug only into a power outlet that has a
u
functioning protective earth connection.
Use only cables and plugs which are in perfect working condition.
u
Connect the laser console only to a power supply that corresponds
u
to the values specified on the rating plate.
Do not use multiple sockets!
u
Do not use extension cables for power supply!
u
Do not touch the system if your body is charged electrostatically
u
and the system is not grounded.
Observe the instructions regarding electromagnetic compatibility
u
(EMC).
Set the system up so that the power cable can be disconnected
u
from the power supply quickly and without any supplementary
means.
Connect the laser console using the power cable intended for use
u
with the laser console. If the laser console is mounted on an
IT 1060.i / IT 760.i instrument table, it will be powered through
this table.
Perform the electrical installation in conformance to
u
IEC 60364-7-710 or the applicable national regulations. This
includes the integration of a ground fault circuit interrupter (GFCI).
Do not touch the patient and the connections of the system simul-
u
taneously.
Live parts are accessible inside the system. If you remove the housing,
you are exposed to the risk of an electric shock.
Never open the system!
u
Documentation set
VISULAS green
000000-2096-101-DokS-en-GB-150322