11 Technical data

11.1 Risk-relevant performance features of clinical
functions
No undesired laser shot and no parameter change in "Treat" step
(READY mode)
No power deviation of more than 20 % without warning message
and no power deviation of more than 50 % without error
message in active shooting mode
Accurate positioning of laser beam
Possibility to stop laser emission in hazardous situations
Protective measures to limit laser radiation in viewing optics

11.2 System classification

The system is classified as follows:
UMDNS No.

11.3 VISULAS green laser system

Laser type
Laser class
Wavelength of therapy beam
Pulse duration (single pulse)
Pulse duration (multi spot)
Pulse duration (selective)
Max. power at cornea
Maximum power on laser exit
aperture of laser source
Maximum power at cornea (cw
mode)
Max. laser energy at cornea
(selective mode)
Pulse interval

11.1 Risk-relevant performance features of clinical functions

17-446 (VISULAS green)
Value
Frequency doubled Nd:YVO
diode-pumped, cw
Class 4 (in accordance with IEC
60825-1)
532 nm
10 ms to 2500 ms, cw (max. 180 s)
10/20/30/40/50 ms
150 ns to 5 µs
1500 mW
(LSL green, VISULINK 532/U,
LIO VISULAS green, laser endoprobe)
3 W (2.5 W + 20 %)
1500 mW
65 µJ (corresponds to a fluence of
2.6 J/cm²)
Adjustable from 10 ms to 6000 ms