Zeiss VISULAS green Manual
Hide thumbs Also See for VISULAS green:
Table of Contents

Advertisement

VISULAS green
Documentation set

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the VISULAS green and is the answer not in the manual?

Questions and answers

Summary of Contents for Zeiss VISULAS green

  • Page 1 VISULAS green Documentation set...
  • Page 2 , ClearView , SpotView , InsightView TouchControl , are either trademarks or registered trademarks of Carl Zeiss Meditec AG or other companies of the ZEISS Group in Germany and/or other countries. All other trademarks in this document are the property of their respective owners.
  • Page 3: Table Of Contents

    VISULAS green version for endocoagulation ................3.1.2 VISULAS green version with LSL green classic laser slit lamp ............3.1.3 VISULAS green version with LSL green comfort laser slit lamp ............. 3.1.4 VISULAS green version with VISULINK 532/U ................Markings on the system ....................System structure......................000000-2096-101-DokS-en-GB-150322 3 / 136...
  • Page 4 Table of Contents Documentation set VISULAS green 3.3.1 System configuration........................3.3.2 System overview.......................... 3.3.3 LSL green laser slit lamp ......................Controls, display elements and connectors..............3.4.1 Laser console..........................3.4.2 Control panel (TouchControl panel) ..................... 3.4.3 LSL green comfort laser slit lamp ....................
  • Page 5 Technical data................... 111 11.1 Risk-relevant performance features of clinical functions ..........111 11.2 System classification......................111 11.3 VISULAS green laser system .................... 111 11.4 LSL green laser slit lamp....................113 11.5 Ambient conditions ......................113 11.6 Electromagnetic compatibility ..................114 11.6.1...
  • Page 6 Table of Contents Documentation set VISULAS green 12.2 Laser endoprobes ......................122 12.3 Connect the door contact isolator to the laser console............ 124 12.4 Physician's safety filter ....................125 Disposal of the system ................127 References ....................129 Glossary ....................131 Index ......................
  • Page 7: Notes On The Instructions For Use

    If you have any questions or comments concerning these instruc- tions for use or the system, please contact ZEISS Service. You can find the ZEISS representative for your country online on the following webpage: www.zeiss.com/med 1.5 Conventions in this document Certain types of information are specially marked in this document for better recognition.
  • Page 8: Conventions In A Course Of Action

    1 Notes on the instructions for use Documentation set 1.6 Applicable documents VISULAS green This is software code or program text. Names of software dialogs, fields or menus, and software messages are marked by quotation marks: • "View" menu. •...
  • Page 9: Safety Notes

    This includes: • Photocoagulation of the retina • Trabeculoplasty for treatment of glaucoma • Iridotomy for treatment of glaucoma VISULAS green is intended for use in Selective Laser Trabeculoplasty (SLT). 000000-2096-101-DokS-en-GB-150322 9 / 136...
  • Page 10: Clinical Benefits

    2 Safety notes Documentation set 2.2 Intended field of application VISULAS green 2.2.2 Clinical benefits Treatment with the VISULAS green photocoagulation laser, including all components and accessories, offers a clinical benefit in the treatment of various eye diseases with the aim of preventing disease progression or averting severe vision loss.
  • Page 11: Indications

    Documentation set 2 Safety notes VISULAS green 2.2 Intended field of application 2.2.4 Indications Photocoagulation • Retinal diseases leading to vascular proliferation due to retinal ischemia, e.g., proliferative diabetic retinopathy, vascular occlusion • Macular edema, e.g. diabetic macula edema •...
  • Page 12 2 Safety notes Documentation set 2.2 Intended field of application VISULAS green • Ocular hypertension • Disorders of choroid and retina (e.g. detachment, hemorrhage, photocoagulation scar) following surgery Iridotomy and laser trabeculoplasty (ALT) • Ocular pain • Visual disturbances (e.g. dysphotopsia, blurred vision) •...
  • Page 13: Responsibilities And Duties Of The Responsible Organization

    To facilitate access for the entire operating personnel: If necessary, request further copies of the instructions for use from ZEISS. Define the required skills for handling the system and provide information on who is authorized for which activities.
  • Page 14 2 Safety notes Documentation set 2.3 Responsibilities and duties of the responsible organization VISULAS green Safety inspections may be carried out only by the manufacturer or persons authorized by the manufacturer. Keep within the prescribed intervals. Perform the inspections to the prescribed extent.
  • Page 15: Notification To Manufacturers And Authorities

    Accessories and additional equipment If you want to connect accessories or additional equipment to the system: Contact your ZEISS contact [} 7]. Additional equipment connected to medical electrical equipment must comply with the corresponding IEC or ISO standards (e.g. IEC 62368-1 for data processing equipment).
  • Page 16: Measures And Duties Of The Operator

    2 Safety notes Documentation set 2.4 Measures and duties of the operator VISULAS green 2.4 Measures and duties of the operator Electrical safety Switch the system off every time before disconnecting it from the power supply, or if you are not going to use the system for any length of time.
  • Page 17 Ensure that no liquids can enter the system. Decommissioning If one of the following events should occur, disconnect the cable from the power supply, label the system clearly as being out of service and report the problem to ZEISS Service: • Electric shocks •...
  • Page 18: Risk From Incoherent Optical Radiation

    8 min (filter position: white). If you have any questions concerning permissible longer exposure times with other filter positions, please contact ZEISS service. When operating the devices, avoid looking directly into the light emitting prism.
  • Page 19: Safety Officer

    A laser warning lamp must be installed at the entrance door indicating when the laser is in operation. The ZEISS laser warning lamp should preferably be used. The supplied remote door interlock switches the laser off automatically when the door to the laser area is opened to gain access. If a laser warning light from a third-party supplier is used, the door interlock may only be used in conjunction with the door contact isolator.
  • Page 20: Laser Safety Goggles / Eye Patches

    • a surgical microscope equipped with a 532 nm physician's safety filter • a ZEISS laser slit lamp or • the LIO VISULAS green laser indirect ophthalmoscope. The responsible organization is responsible for providing proper laser safety goggles. The safety goggles must be approved for the wavelength of the laser and for the type and intensity of the laser radiation in your country.
  • Page 21: Danger Caused By Direct Irradiation And Reflection

    Documentation set 2 Safety notes VISULAS green 2.7 Risk of fire 2.6.6 Danger caused by direct irradiation and reflection The system and its applicators emit visible laser radiation which is dangerous to the human eye and skin. As an application device the LIO VISULAS green laser indirect ophthal- moscope may only be operated together with the VISULAS green laser...
  • Page 22: Maintenance Measures

    Maintenance procedures (maintenance and repairs) which are not specified in these instructions for use may only be carried out by persons authorized by Carl Zeiss Meditec and solely according to the service instructions issued by Carl Zeiss Meditec. For planning and implementing these maintenance and care procedures please contact ZEISS Service or your local dealer.
  • Page 23: Description Of The Device

    Documentation set 3 Description of the device VISULAS green 3.1 Package check list 3 Description of the device 3.1 Package check list 3.1.1 VISULAS green version for endocoagulation • VISULAS green (laser console with control panel) • Standard foot switch (1.6 m) •...
  • Page 24: Visulas Green Version With Visulink 532/U

    3.1 Package check list VISULAS green • Cap for fixation lamp mounting • Power supply cable (2.5 m) • Documentation set 3.1.4 VISULAS green version with VISULINK 532/U • VISULAS green • VISULINK 532/U (2.0 m) • Slit lamp adapter • Standard foot switch (1.6 m) •...
  • Page 25: Markings On The System

    Documentation set 3 Description of the device VISULAS green 3.2 Markings on the system 3.2 Markings on the system The label texts on the VISULINK 532/U, the LIO VISULAS green laser indirect ophthalmoscope, the laser endoprobes, the SL cam compact, the SL Workstation and the instrument tables are described in the corresponding instructions for use.
  • Page 26 3 Description of the device Documentation set 3.2 Markings on the system VISULAS green Labels Explanation Identification label with unique device identifi- cation code (Unique device identification label) for laser console with single/dual fiber port Catalog number / part number...
  • Page 27 Documentation set 3 Description of the device VISULAS green 3.2 Markings on the system Control panel (TouchControl panel) Symbol Name Identification label Catalog number / part number Serial number LSL green comfort/classic laser slit lamp Symbol Name Type label Manufacturer Date of manufacture...
  • Page 28 3 Description of the device Documentation set 3.2 Markings on the system VISULAS green Symbol Name "Laser beam" warning label "Laser aperture" warning label Standard / TouchControl foot switch Symbol Name Type label Manufacturer IPX8 Ingress protection rating (protection against continuous immersion in water)
  • Page 29: System Structure

    Documentation set 3 Description of the device VISULAS green 3.3 System structure 3.3 System structure 3.3.1 System configuration Slit lamp adapters SL 20 / SL 30 SL 220 LIO VISULAS green laser indirect SL 120 / SL 130 ophthalmoscope VISULINK 532/U SL 115 Classic 900 BM / 900 BQ VISULAS green laser console...
  • Page 30: System Overview

    3 Description of the device Documentation set 3.3 System structure VISULAS green 3.3.2 System overview Figure 2: System overview (version shown is the VISULAS green comfort) Laser slit lamp Control panel Instrument table Laser console Table mounting for laser console Foot switch...
  • Page 31: Lsl Green Laser Slit Lamp

    Documentation set 3 Description of the device VISULAS green 3.3 System structure 3.3.3 LSL green laser slit lamp Figure 3: Laser slit lamp (version shown is the LSL green comfort) Eyepiece with adjustable eye cup Binocular tube Microscope stand with fixed physician’s...
  • Page 32 3 Description of the device Documentation set 3.3 System structure VISULAS green 3.3.3.1 Beam delivery in the laser slit lamp The laser beam is interlaced into the illumination beam path, so that the slit image and the laser spot can move around the fundus in synchrony.
  • Page 33: Controls, Display Elements And Connectors

    Documentation set 3 Description of the device VISULAS green 3.4 Controls, display elements and connectors 3.4 Controls, display elements and connectors 3.4.1 Laser console The laser console is the central power supply and control unit for the VISULAS green. Figure 5: Control elements, displays and ports on the laser console (version shown is the console version with...
  • Page 34 3 Description of the device Documentation set 3.4 Controls, display elements and connectors VISULAS green Item Icon/Name Explanation Optical fiber for LIO VISULAS green laser indirect ophthalmo- (laser console with dual fiber scope port) LED ring on the corresponding fiber optic port lights up when selected •...
  • Page 35: Control Panel (Touchcontrol Panel)

    Documentation set 3 Description of the device VISULAS green 3.4 Controls, display elements and connectors 3.4.2 Control panel (TouchControl panel) The laser console is operated via touch screen menus or using the combined rotary knob / button on the control panel.
  • Page 36 3 Description of the device Documentation set 3.4 Controls, display elements and connectors VISULAS green 3.4.2.1 Changing parameter values on the touch screen Item Icon/ Explanation Name Parameter Parameter to be changed. The parameters are activated by tapping the touch screen.
  • Page 37: Lsl Green Comfort Laser Slit Lamp

    Documentation set 3 Description of the device VISULAS green 3.4 Controls, display elements and connectors 3.4.3 LSL green comfort laser slit lamp Figure 7: Control elements on the LSL green comfort laser slit lamp Item Name Explanation Height adjustment for chin rest The optimum eye level for the patient can be set by turning the height adjustment of the chin rest.
  • Page 38 3 Description of the device Documentation set 3.4 Controls, display elements and connectors VISULAS green Item Name Explanation Filter selector White line pointing up: Full aperture combined with a heat- absorbing filter. White line to the patient: Blue filter for tonometry and fluores- cence observations.
  • Page 39 Documentation set 3 Description of the device VISULAS green 3.4 Controls, display elements and connectors 3.4.3.1 LSL InsightView display and TouchControl LSL center switch In the LSL green comfort laser slit lamp, the current parameter values can be seen on the integral LSL InsightView display in the observer's field of vision.
  • Page 40 3 Description of the device Documentation set 3.4 Controls, display elements and connectors VISULAS green Icon / Name Explanation Adjustable parameters • Laser power • Pulse duration • Pause between laser pulses • Brightness of the LSL InsightView display (LSL green comfort only) •...
  • Page 41: Lsl Green Classic Laser Slit Lamp

    Documentation set 3 Description of the device VISULAS green 3.4 Controls, display elements and connectors 3.4.4 LSL green classic laser slit lamp Figure 9: Control elements on the LSL green classic laser slit lamp 000000-2096-101-DokS-en-GB-150322 41 / 136...
  • Page 42 3 Description of the device Documentation set 3.4 Controls, display elements and connectors VISULAS green Item Icon/Name Explanation Height adjustment for chin rest The optimal eye level for the patient can be set by turning the height adjustment of the chin rest.
  • Page 43: Standard Foot Switch

    Documentation set 3 Description of the device VISULAS green 3.4 Controls, display elements and connectors 3.4.5 Standard foot switch Figure 10: Standard foot switch Item Name Explanation Protective cover Trigger Depending on the setting, pressing the trigger will emit one or more laser pulses which are applied to the patient's eye.
  • Page 44: Touchcontrol Foot Switch

    3 Description of the device Documentation set 3.4 Controls, display elements and connectors VISULAS green 3.4.6 TouchControl foot switch Figure 11: TouchControl foot switch Item Name Explanation Lever for changing This lever is used for changing the values of parameters the selected parameter. Tapping the left side of the lever reduces the parameter value;...
  • Page 45: Functional Description

    Documentation set 3 Description of the device VISULAS green 3.5 Functional description 3.5 Functional description Pumped by a diode laser at 810 nm, a laser beam with a wavelength of 1064 nm, the frequency of which is doubled by means of a crystal, is produced in a solid state medium.
  • Page 46: Safety Systems

    3 Description of the device Documentation set 3.5 Functional description VISULAS green 3.5.1 Safety systems Safety system Effect Plug-type connector for The system incorporates a plug-type remote-controlled locking connector for remote-controlled locking (e.g. door contact). Laser emission is inter- rupted when the contacts of the plug-type connector are open.
  • Page 47: Installation

    Installation and operating errors may result in personal injury and material damage. The system may only be installed and operated for the first time by persons who have been authorized by Carl Zeiss Meditec. Electrical hazard WARNING! The use of unsuitable power cables could result in electric shock.
  • Page 48: Connecting

    4 Installation Documentation set 4.2 Connecting VISULAS green Mechanical hazard due to instability CAUTION! Injury may be caused by falling parts. When selecting a suitable table, ensure that the combination of table and device is stable up to a tilt angle of 10°. The table must be designed for 4 times the weight of the device configuration.
  • Page 49: Mounting The Foot Switch On The Foot Of The Instrument Table

    Documentation set 4 Installation VISULAS green 4.3 Mounting the foot switch on the foot of the instrument table 4.3 Mounting the foot switch on the foot of the instrument table To fasten the TouchControl foot switch to the corrugated feet of the IT 760.i / IT 1060.i instrument tables, you must first remove the base...
  • Page 50: Moving The Instrument Table With The Devices

    4 Installation Documentation set 4.4 Moving the instrument table with the devices VISULAS green Action 1. Use a screwdriver to remove the grub screw from the hole in which the foot switch is to be secured. The adapter plate for the foot switch cannot be removed. The upper plate on which the foot switch is mounted can be rotated for ergonomic adjustment of the foot switch.
  • Page 51: Daily Startup

    Documentation set 5 Daily startup VISULAS green 5.1 Preparation safety 5 Daily startup 5.1 Preparation safety Prior to using the system, the user must ensure that it is in a good condition and fully functioning. Furthermore, the user must follow the instructions in the instructions for use.
  • Page 52 Empty page, for your notes...
  • Page 53: Operation

    Documentation set 6 Operation VISULAS green 6.1 Operation safety 6 Operation 6.1 Operation safety Hazard due to laser radiation WARNING! If the VISULAS green is used in a different manner to that specified in these instructions, this may result in dangerous radiation exposure.
  • Page 54 6 Operation Documentation set 6.1 Operation safety VISULAS green Hazard due to laser radiation WARNING! Using an unsuitable contact lens for laser treatment can result in injuries caused by laser radiation. Always use the correct contact lens for laser treatment.
  • Page 55: Operation With Laser Slit-Lamp

    Documentation set 6 Operation VISULAS green 6.2 Operation with laser slit-lamp Hazard due to laser radiation WARNING! The use of the SLT option on the retina may result in injury from laser radiation. Do not use the SLT option on the retina.
  • Page 56: Positioning The Contact Lenses

    6 Operation Documentation set 6.2 Operation with laser slit-lamp VISULAS green 9. Select the other magnification steps on the slit lamp. ð The image should remain sharply in focus at all magnifications. If not, the procedure should be repeated. 10. Make a note of the eyepiece settings for future treatment sessions.
  • Page 57: Operation With Laser Endoprobes

    Documentation set 6 Operation VISULAS green 6.3 Operation with laser endoprobes 4. Switch to "Treat" work step. 5. Actuate the manual or foot trigger to release laser pulses according to the parameter setting (see Section Triggering the laser). NOTE! On the navigation bar select the "Prepare" work step, if no laser treatment is to be performed.
  • Page 58: Laser Treatment With The Laser Endoprobe

    6 Operation Documentation set 6.3 Operation with laser endoprobes VISULAS green 6.3.2 Laser treatment with the laser endoprobe Hazard due to laser radiation WARNING! Imprecise localization of the target area may result in injury. Only operate the foot switch if you have clearly localized the target area and are sure that you wish to use the therapy laser beam.
  • Page 59: Switching Off The System

    Documentation set 6 Operation VISULAS green 6.4 Switching off the system 11. Localize the target area and adjust the size of the laser spot. The tip of the laser endoprobe should normally be 2 mm to 5 mm from the retina, depending on the spot size and flux density required.
  • Page 60 Empty page, for your notes...
  • Page 61: Software Description

    Documentation set 7 Software description VISULAS green 7.1 Data security 7 Software description 7.1 Data security • VISULAS green can export data directly to a USB drive, for example treatment parameters as PDF documents or log files. • VISULAS green in combination with the "Laser Treatment Workplace"...
  • Page 62: Overview

    7 Software description Documentation set 7.2 Overview VISULAS green 7.2 Overview Upon switching the system on, the user interface is displayed on the control panel. Figure 15: User interface on control panel Item Icon/Name Explanation Settings Set brightness and volume, open the “Settings” dialog window.
  • Page 63: On-Screen Keyboard

    Documentation set 7 Software description VISULAS green 7.3 On-screen keyboard 7.3 On-screen keyboard Data is entered using the on-screen keyboard. It appears automatically as soon as the cursor is inserted into a text box. Figure 16: On-screen keyboard 7.4 "Configure" work step In the first "Configure"...
  • Page 64 7 Software description Documentation set 7.4 "Configure" work step VISULAS green Item Icon/Name Explanation Users Select a user. Therapy Select a therapy. Applicator Select an applicator. The LIO and ENDO applicators are only available if Retina therapy has been selected.
  • Page 65: Prepare" Work Step

    Documentation set 7 Software description VISULAS green 7.5 "Prepare" work step 7.5 "Prepare" work step In the "Prepare" work step set the laser parameters for the current treatment. 7.5.1 Single-spot mode In Single-spot mode the laser is released by the manual or foot trigger.
  • Page 66 7 Software description Documentation set 7.5 "Prepare" work step VISULAS green Item Icon/Name Explanation Single-spot mode is selected. Single-spot Select Multi-spot mode. The Multi-spot mode is only available if the slit lamp has been selected as applicator. Multi-spot Set the interval between two laser pulses. The pause can be adjusted using the rotary knob, tapping the two arrow keys or with a circular motion on the displayed arc.
  • Page 67: Multi-Spot Mode (Optional)

    Documentation set 7 Software description VISULAS green 7.5 "Prepare" work step Item Icon/Name Explanation Applicator Displays the selected applicator. Slit lamp has been selected as the applicator. LIO VISULAS green laser indirect ophthalmoscope has been selected as the applicator. Laser endoprobe has been selected as the applicator.
  • Page 68 7 Software description Documentation set 7.5 "Prepare" work step VISULAS green Item Icon/Name Explanation Set the laser pulse length. The pulse length can be adjusted using the rotary knob, tapping the two arrow keys or with a circular motion on the displayed arc.
  • Page 69 Documentation set 7 Software description VISULAS green 7.5 "Prepare" work step Item Icon/Name Explanation Applicator Displays the selected applicator. Slit lamp has been selected as the applicator. 7.5.2.1 Configure the multi-spot sequence Figure 20: Configure multi-spot sequence. Item Icon/Name Explanation Arrange the laser spots in a line. A single linear spot sequence will be triggered.
  • Page 70 7 Software description Documentation set 7.5 "Prepare" work step VISULAS green Item Icon/Name Explanation Preview of the multi-spot Displays the multi-spot sequence with the selected settings. sequence Reduce the laser spot spacing. Alternatively, the distance between the spots may be decreased.
  • Page 71: Selective Mode (Optional)

    Documentation set 7 Software description VISULAS green 7.5 "Prepare" work step 7.5.3 Selective mode (optional) Selective mode is only available if you have the appropriate license. Select the "Glaucoma" therapy and the "Selective (SLT)" treatment in the "Configure" work step. In the "Selective (SLT)" treatment, the laser is always triggered by the manual trigger.
  • Page 72 7 Software description Documentation set 7.5 "Prepare" work step VISULAS green Item Icon/Name Explanation Reset the laser spot counter to zero. Reset Displays the energy density (laser fluence) of the laser beam at the area to be treated. The energy density is a measure of the therapeutic effect on the eye.
  • Page 73 Documentation set 7 Software description VISULAS green 7.5 "Prepare" work step The following Scheie grading system and table show recommended starting values for SLT treatment. The stronger the pigmentation of the patient's eye, the lower the energy which should be selected.
  • Page 74 7 Software description Documentation set 7.5 "Prepare" work step VISULAS green In the following table you will find a conversion table of laser energies of VISULAS lesions consisting of 52 single spots to corresponding Nd:YAG single shot lesions at a diameter of 400 μm.
  • Page 75: Treat" Work Step

    Documentation set 7 Software description VISULAS green 7.6 “Treat” work step 7.6 “Treat” work step In the "Treat" work step, information on the current treatment is displayed. The laser is active. Once all parameters have been set, the laser beam can be released with the manual or foot trigger.
  • Page 76 7 Software description Documentation set 7.6 “Treat” work step VISULAS green Figure 23: "Treat" work step in single spot mode Item Icon/Name Explanation Set the laser power. The laser power can be changed by using the rotary knob, by tapping the two arrow keys or by circular gestures on the displayed arc.
  • Page 77 Documentation set 7 Software description VISULAS green 7.6 “Treat” work step Item Icon/Name Explanation Set the pause between two laser pulses. The pause duration can be changed using the rotary knob, by tapping the two arrow keys or by circular gestures on the displayed arc.
  • Page 78: Multi-Spot Mode (Optional)

    7 Software description Documentation set 7.6 “Treat” work step VISULAS green 7.6.2 Multi-spot mode (optional) Multi spot mode is only available if you have the multi spot "VISULAS VITE 2.0" software license and one of the LSL green comfort or LSL green combi laser slit lamp models, and if the laser slit lamp is selected as applicator.
  • Page 79 Documentation set 7 Software description VISULAS green 7.6 “Treat” work step Figure 24: "Treat" work step in multi spot mode Item Icon/Name Explanation Set the laser power. The laser power can be changed by using the rotary knob, by tapping the two arrow keys or by circular gestures on the displayed arc.
  • Page 80: Selective Mode (Optional)

    7 Software description Documentation set 7.6 “Treat” work step VISULAS green Item Icon/Name Explanation Reset the laser spot counter to zero. Reset Displays the energy density (laser fluence) of the laser beam at the area to be treated. The energy density is a measure of the therapeutic effect on the eye.
  • Page 81 Documentation set 7 Software description VISULAS green 7.6 “Treat” work step To emit the laser beam, hold the trigger for about 0.5 seconds. As a result, the SLT lesion, consisting of 52 individual spots, is completely applied. NOTE! If you use a contact lens with deflection mirrors, be sure to keep the laser beam away from the mirrored edges.
  • Page 82 7 Software description Documentation set 7.6 “Treat” work step VISULAS green Figure 25: "Prepare" step for the "Selective (SLT)" treatment Item Icon/Name Explanation Select the contact lens (e.g. Latina Gonio) using the drop-down list. Use only contact lenses with a magnification factor of 1.0,...
  • Page 83: Document" Work Step

    Documentation set 7 Software description VISULAS green 7.7 "Document" work step Item Icon/Name Explanation Displays the number of laser pulses triggered, showing the number of individual laser spots and, in parentheses, the number of 400 µm lesions applied. Laser pulse counter Displays the set laser spot size when no contact lens is selected.
  • Page 84 7 Software description Documentation set 7.7 "Document" work step VISULAS green NOTE! After exiting the "Document" step, the treatment para- meters and the report are no longer available on the VISULAS green. Figure 26: Treatment report - Retina therapy Figure 27: : Treatment report - glaucoma therapy (selective trabeculoplasty)
  • Page 85: Settings

    Documentation set 7 Software description VISULAS green 7.8 Settings 7.8 Settings The "Settings" dialog window can be opened by tapping the "Settings" icon in the top right corner of the user interface. Figure 28: “Settings” menu Item Icon/Name Explanation Access and adjust the settings on the user interface.
  • Page 86: General

    7 Software description Documentation set 7.8 Settings VISULAS green 7.8.1 General General settings on the user interface such as language or appearance can be adjusted in this window. Figure 29: "General” dialog area Item Icon/Name Explanation Language Select a user interface language using the drop-down list.
  • Page 87: Treatments

    Documentation set 7 Software description VISULAS green 7.8 Settings 7.8.2 Treatments This window can be used to pre-set frequently used treatment config- urations. The latter are linked to user profiles and become available when the respective user is selected. Figure 30: “Treatments” dialog area...
  • Page 88 7 Software description Documentation set 7.8 Settings VISULAS green 7.8.2.1 User management This window is used to manage users. Profiles may be added, edited or deleted. Figure 31: Overview of current users Item Icon/Name Explanation Delete Delete selected user. The action must be confirmed in the following dialog window.
  • Page 89 Documentation set 7 Software description VISULAS green 7.8 Settings Add or edit user Figure 32: “User” dialog window Item Icon/Name Explanation Input area Enter user data. Cancel Close the dialog window without saving. Save Save and close dialog window. 7.8.2.2 Retina therapy Figure 33: “Select the treatment mode.”...
  • Page 90 7 Software description Documentation set 7.8 Settings VISULAS green 7.8.2.2.1 Single-spot treatments In this dialog, you can manage treatment configurations. Treatments may be added, edited or deleted. Figure 34: Overview of currently available single-spot treatments Item Icon/Name Explanation Name Name of treatment...
  • Page 91 Documentation set 7 Software description VISULAS green 7.8 Settings Add or edit a treatment Figure 35: Dialog for adding or editing a treatment Item Icon/Name Explanation Name Name of treatment Set the laser power. Power Set the pulse duration. Pulse Set the pause between two laser pulses.
  • Page 92 7 Software description Documentation set 7.8 Settings VISULAS green 7.8.2.2.2 Multi-spot treatments In this dialog, you can manage treatment configurations. Treatments may be added, edited or deleted. Figure 36: Overview on currently available multi spot treatments Item Icon/Name Explanation Name Name of treatment...
  • Page 93 Documentation set 7 Software description VISULAS green 7.8 Settings Item Icon/Name Explanation Recommended laser spot size Spot size Close Close the dialog. Add a treatment. Edit Edit the selected treatment. Delete Delete the selected treatment. The action must be confirmed in the following dialog window.
  • Page 94: Contact Lenses

    7 Software description Documentation set 7.8 Settings VISULAS green Item Icon/Name Explanation Multi spot sequence Configure th multi spot sequence. Save Save the settings. Cancel Close the dialog window without saving. 7.8.2.3 Glaucoma therapy Glaucoma therapy is performed in single spot mode or selective mode.
  • Page 95 Documentation set 7 Software description VISULAS green 7.8 Settings Add or edit a contact lens profile Figure 39: Dialog window for adding or editing a contact lens Item Icon/Name Explanation Name Enter or edit the name of a contact lens. Magnification Set the laser spot magnification factor using the [+] and [-] buttons.
  • Page 96: System Information And Demo Mode

    7 Software description Documentation set 7.8 Settings VISULAS green 7.8.4 System information and DEMO mode This window is used to view information on the product version and to activate/deactivate the DEMO mode. Figure 40: “System” dialog area Item Icon/Name Explanation Product Display of the product name.
  • Page 97 Documentation set 7 Software description VISULAS green 7.8 Settings Details and DEMO mode Figure 41: “Details” dialog window Item Icon/Name Explanation Information area Displays details of the product version. Close Close the window. DEMO Activate / deactivate DEMO mode. The action must be confirmed in the following dialog window.
  • Page 98: Maintenance

    7 Software description Documentation set 7.8 Settings VISULAS green 7.8.5 Maintenance This window contains license information. Service-relevant actions can be triggered here. Figure 42: “Maintenance” dialog area Item Icon/Name Explanation Export log files The [Log file export] button opens a browser window in which the export directory can be selected and the export of log files is confirmed.
  • Page 99: Cleaning And Disinfection

    Documentation set 8 Cleaning and disinfection VISULAS green 8.1 Safety when cleaning and disinfecting the device 8 Cleaning and disinfection 8.1 Safety when cleaning and disinfecting the device Electrical hazard due to penetration of moisture WARNING! The penetration of moisture into the devices may cause an electric shock.
  • Page 100: Cleaning

    8 Cleaning and disinfection Documentation set 8.2 Cleaning VISULAS green 8.2 Cleaning Action Do not use acetone and acetone-based cleaners to clean the devices. ð These cleaning agents may damage the surfaces of the devices. All parts of the device casing may be wiped down with a moist but not drip-wet cloth.
  • Page 101: Maintenance

    These must be adhered Unless otherwise regulated by local legal provisions: Have preventive maintenance carried out on your device by ZEISS Service or have an electrical safety inspection according to IEC 62353 performed by a qualified technician annually. When performing the inspection, observe the service instructions issued by Carl Zeiss Meditec and/or the following instructions.
  • Page 102 9 Maintenance Documentation set 9.1 Testing electrical safety VISULAS green Figure 44: Measuring points on laser slit lamp 102 / 136 000000-2096-101-DokS-en-GB-150322...
  • Page 103 Documentation set 9 Maintenance VISULAS green 9.1 Testing electrical safety Figure 45: Measuring points on instrument table 3. If there are doubts about the efficacy of the insulation (such as multiple triggering of the fault circuit interrupter or other protective devices at the place of use, or traces of liquid on the system which suggest the penetration of liquid), measure the insulation resistance with a test voltage of 500 V.
  • Page 104: Perform Safety Inspections

    Regular calibration of the laser system is not required by the user. The safety features and equipment may only be inspected by persons authorized to do so by Carl Zeiss Meditec. When performing the inspection, observe the service instructions issued by Carl Zeiss Meditec.
  • Page 105: Replacing The Halogen Bulb In The Laser Slit Lamp

    Documentation set 9 Maintenance VISULAS green 9.3 Replacing the halogen bulb in the laser slit lamp 9.3 Replacing the halogen bulb in the laser slit lamp Danger of exposure to the hot halogen bulb CAUTION! Touching the hot halogen bulb may result in minor burn injuries.
  • Page 106: Replacing The Fuse

    9 Maintenance Documentation set 9.4 Replacing the fuse VISULAS green 9.4 Replacing the fuse Damage caused by incorrect fuse NOTE Inserting an incorrect fuse may damage the system. Disconnect from the power supply before changing the fuse! Only use fuses which meet the specifications given on the sign containing the fuse markings.
  • Page 107: Replacing The Light-Emitting Diode (Led) Of The Fixation Lamp (Optional)

    Documentation set 9 Maintenance VISULAS green 9.5 Replacing the light-emitting diode (LED) of the fixation lamp (optional) 9.5 Replacing the light-emitting diode (LED) of the fixation lamp (optional) Damage caused by unsuitable LED NOTE Installing an unsuitable LED may cause the fixation lamp to overheat and be destroyed.
  • Page 108 Empty page, for your notes...
  • Page 109: Troubleshooting

    Warning message: Remedy the fault using the troubleshooting tables below. If a fault occurs, which cannot be remedied by using the following tables, please contact ZEISS Service, citing the error number displayed. Label the system as being out of order, remove the key from the key switch and disconnect the power plug.
  • Page 110: Faults (No Messages)

    "Settings / System / work step. The laser cannot be Information" to deactivate the triggered. DEMO mode. 10.3 Service information You can find the ZEISS representative for your country online on the following webpage: www.zeiss.com/med 110 / 136 000000-2096-101-DokS-en-GB-150322...
  • Page 111: Technical Data

    • Protective measures to limit laser radiation in viewing optics 11.2 System classification The system is classified as follows: UMDNS No. 17-446 (VISULAS green) 11.3 VISULAS green laser system Value Laser type Frequency doubled Nd:YVO diode-pumped, cw Laser class Class 4 (in accordance with IEC...
  • Page 112 11 Technical data Documentation set 11.3 VISULAS green laser system VISULAS green Value Aiming beam Diode, 620 nm to 650 nm, max. 1 mW at the cornea Aiming beam laser class Complies with the class 2 threshold value according to IEC 60825-1 Rated voltage and frequency 100 V to 240 V AC (±10 %);...
  • Page 113: Lsl Green Laser Slit Lamp

    Documentation set 11 Technical data VISULAS green 11.4 LSL green laser slit lamp 11.4 LSL green laser slit lamp LSL green comfort LSL green classic Device group 2 in accordance with ISO 15004-2 Laser treatment Continuously adjustable from 50 µm to 1000 µm...
  • Page 114: Electromagnetic Compatibility

    If it is nonetheless necessary to operate VISULAS green in the aforementioned manner, observe the VISULAS green and the other devices to ensure they are operating correctly. Do not use portable HF communications equipment (including peripheral devices such as antenna cable and external antennas) within a radius of 30 cm around the VISULAS green, including...
  • Page 115 CAUTION! The use of accessories, all types of transducers and cables not specified in these instructions for use or not sold by Carl Zeiss Meditec as replacement parts may result in higher electromagnetic emissions or reduced immunity of the system and thus in faulty operation.
  • Page 116: Guidelines And Manufacturer's Electromagnetic Compatibility Declaration

    11 Technical data Documentation set 11.6 Electromagnetic compatibility VISULAS green 11.6.3 Guidelines and manufacturer's electromagnetic compatibility declaration The following guideline applies exclusively to accessories specified for and delivered with the device by the manufacturer or seller. Guidance and manufacturer's declaration – electromagnetic emissions The VISULAS green is intended for operation in an electromagnetic environment as specified below.
  • Page 117 Documentation set 11 Technical data VISULAS green 11.6 Electromagnetic compatibility Guidance and manufacturer’s declaration - electromagnetic immunity The VISULAS green is intended for use in the electromagnetic environment specified below. The customer or the user of the VISULAS green should ensure that it is used in such an environment.
  • Page 118 11 Technical data Documentation set 11.6 Electromagnetic compatibility VISULAS green Guidance and manufacturer’s declaration – electromagnetic immunity The VISULAS green is intended for use in the electromagnetic environment specified below. The customer or the user of the VISULAS green should ensure that it is used in such an environment.
  • Page 119: Optional Accessories And Spare Parts

    Convergent tube f = 140 mm • VISULINK 532/U (2.0 m) (see instructions for use of VISULINK 532/ • 532 adapter for ZEISS SL 120 / SL 130 • 532 adapter for ZEISS SL 115 Classic • 532 adapter for ZEISS 20 SL / 30 SL •...
  • Page 120 An up-to-date list of accessories approved for your country can be obtained from your ZEISS contact. You can find the ZEISS representative for your country online on the following webpage: www.zeiss.com/med Only use accessories and spare sparts which are approved by ZEISS for this system.
  • Page 121: Instrument Table

    Documentation set 12 Optional accessories and spare parts VISULAS green 12.1 Instrument table 12.1 Instrument table Electrical hazard due to improper electrical connection WARNING! Improper electrical installation may cause an electric shock. Connect the laser console via the IT 1060.i / IT 760.i instrument table.
  • Page 122: Laser Endoprobes

    Use the laser endoscope only once. Resterilization is not permitted. It is recommended to use VISULAS green with ZEISS laser endoprobes as the efficacy and safety of this combination has been demonstrated. As a rule, VISULAS green is prepared for use with commercial endoprobes from third party suppliers which have been approved for intraocular treatment.
  • Page 123 Light transmission of laser endoprobes may vary from supplier to supplier. If you change your laser endoprobe supplier, make sure that your local ZEISS service engineer measures and stores the transmission of the new laser endoprobe in the system. This will enable the proper output power indicated on the graphical user interface of the control panel.
  • Page 124: Connect The Door Contact Isolator To The Laser Console

    NOTE! The connection must be installed with strain relief to prevent the plug from being inadvertently pulled out of the socket. NOTE! The door contact of the ZEISS laser warning lamp may be connected without a door contact isolator directly to the interlock socket for door contact.
  • Page 125: Physician's Safety Filter

    Documentation set 12 Optional accessories and spare parts VISULAS green 12.4 Physician's safety filter 12.4 Physician's safety filter Hazard due to laser radiation CAUTION! During laser treatment with a surgical microscope, the physician's eyes may be harmed by laser radiation.
  • Page 126 12 Optional accessories and spare parts Documentation set 12.4 Physician's safety filter VISULAS green • On OPMI CS surgical microscopes, the beam splitter is built into the body of the microscope. It is therefore not possible to mount the physician's safety filter upstream of the beam splitter. If you...
  • Page 127: Disposal Of The System

    For further information on disposing of the system, contact the ZEISS contact person for your country. You can find the ZEISS representative for your country online on the following webpage: www.zeiss.com/med If you resell the system or its components: Inform the buyer that the system is to be disposed of in accordance with the currently applicable regulations.
  • Page 128 Empty page, for your notes...
  • Page 129: References

    Documentation set 14 References VISULAS green ​ 14 References [1] Schargus M, Schrader WF: Retinale Photokoagulation: Techniken - Indikationen - Anwendungen. Z. Prakt. Augenheilkd. 2010, 31, 21-32. [2] Alasil T, Waheed NK (2014) Pan retinal photocoagulation for prolif- erative diabetic retinopathy: pattern scan laser versus argon laser Current opinion in ophthalmology 25: 164-170.
  • Page 130 Empty page, for your notes...
  • Page 131: Glossary

    Documentation set Glossary VISULAS green Glossary Argon laser trabeculoplasty Selective laser trabeculoplasty D (unit of measurement of the refractive Plug-type connector for fiber optic cables power of optical systems) Electromagnetic Compatibility ENDO Laser endoprobe Light emitting diode Laser indirect ophthalmoscope...
  • Page 132 Empty page, for your notes...
  • Page 133: Index

    Documentation set Index VISULAS green Index TouchControl ......Accessories ......... 15, 120 Adjusting the slit lamp ......Instrument table ........50, 121 Ambient conditions ......Intended use/Indication for use.... Beam delivery ........Key switch ........... Changes to the product ......
  • Page 134 Index Documentation set VISULAS green Maintenance ......System ........Treatments........ System overview......... TouchControl panel ......Transport..........Treat........... User profile......... 134 / 136 000000-2096-101-DokS-en-GB-150322...
  • Page 136 Carl Zeiss Meditec AG Goeschwitzer Strasse 51-52 07745 Jena Germany Internet: www.zeiss.com/med Email: info.meditec@zeiss.com 000000-2096-101-DokS-en-GB-150322...