Philips PM3320 Operating Manual page 55

250 ms/s dual channel digital storage oscilloscope
Table of Contents

Advertisement

4-35E
8
EXECUTE
After pushing this softkey, the selected trigger delay value is entered and an AUTO
RETURN
is performed to menu TRIGGER
DELAY.
If after CLEAR the softkey EXECUTE is pressed the trigger delay value keeps its
previous value in the TRIGGER DELAY menu.
For TIME and DiV a message
TRIGGER
DELAY
number out of range.
can be displayed if the entered value exceeds the given range.
For EVENTS
a message
EVENT number out of range.
can be displayed if the entered value exceeds the given range.
UP
Pushing softkey UP increments the trigger delay.
The TIME increases in steps of the time-base setting multiplied by 1 or by 10 depending on
the setting SCREEN or DIV.
DIV increases in steps of 1 division and SCREEN increases in steps of 1 screen. EVENTS.
always increases in steps of 1 event.
DOWN
Pushing softkey DOWN decrements the trigger delay. The TIME decreases in steps of the
time-base setting multiplied by 1 or by 10 depending on the setting SCREEN or DIV.
DIV decreases in steps of 1 divisionand SCREEN decreases in steps of 1 screen. EVENTS,
always decreases in steps of 1 event.
SCREEN
DIV
The text SCREEN
DIV is displayed when TIME or DIV is active.
If the text SCREEN is intensified, the UP and DOWN softkeys operate with 1 screen,
If the text DIV is intensified, the UP and DOWN
softkeys operate with 1 division.
Pushing this softkey changes the selection from SCREEN
to DtV or reverse,
lf EVENTS triggering is operative, this text is not displayed.