Use of probes
The 10:1 passive probes must be properly compensated before use to avoid pulse distortion or
amplitude errors at high frequencies.
For correct adjustment, refer to section 8.1.1.4.
Vertical
As channels A and B are identical, only the procedure for channel A is described.
Vertical mode
— Press the green pushbutton AUTO; a square wave with a vertical amplitude of a few divisions
becomes visible on the screen.
- Press pushbutton VERTICAL MODE. Check that channel A is selected.
- Press softkey A and B. Channel B is visible as a straight line.
-
Press pushbutton VERTICAL
COUPLING.
— Press softkey GROUND
of COUPLING
B.
-
Press pushbutton VERTICAL MODE.
-
Press softkey ADD; only the square wave is visible.
Press softkey A. Check the square wave for minimum jump in vertical direction (< 0,3 div.).
-
Press softkey A-INVERT; the square wave is inverted.
-
Press softkey A-INVERT; the square wave is inverted again (normal).
~ Press softkey PROCESSING; the VERTICAL PROCESSING
menu is visible.
- Press softkey MIN / MAX. Check that the straight horizontal parts of the square wave become
more frayed.
- Press softkey MIN / MAX again; MIN / MAX is off.
- Press softkey RETURN; the VERTICAL
MODE menu
appears.
Vertical coupling
- Press pushbutton VERTICAL COUPLING. Check that AC coupling is selected.
- Press softkey AC DC of VERTICAL COUPLING A; DC is selected.
~
Check
that the displayed square
wave
is shifted upwards
with zero
line on vertical
midposition.
Press softkey GROUND of VERTICAL COUPLING A. Check that the input signal is switched
off, a straight line is visible on the screen.
~
Press softkey AC. DC of channel A; DC is selected and the square wave is visible again.
-
Press softkey OFFSET of channel A.
-
Press softkey UP and DOWN
and check if the signal moves
up and down.
Vertical controls
Press UP/DOWN control AMPL/DIV on it's left side. Check that the amplitude of the square
wave on the screen decreases.
Press UP/DOWN
control AMPL/DIV on it's right side. Check that the amplitude increases.
- Turn VARIABLE left. Check that the amplitude decreases and pilot lamp UNCAL lights up.
- Turn VARIABLE right. Check that the amplitude increases and pilotiamp UNCAL extinguishes.
- Turn SHIFT to the right to move the signal up and to the left to move it down,