Philips PM3320 Operating Manual page 36

250 ms/s dual channel digital storage oscilloscope
Table of Contents

Advertisement

4-16E
7
PROCESSING>
Alter selecting PROCESSING>, the VERTICAL PROCESSING menu
is displayed and input signal processing can be selected.
The text MIN / MAX is not visible in the time base positions faster than
5 us/div.
VERTICAL
PROCESSING
The text AVERG
OFF
is only visibie if average
is selected.
MIN / MAX
4
BW LIMIT
2
AVERG OFF
3
AVERAGE
>
4
o=8
=
\
oe
\
ea
7
RETURN
8
7
1
MIN / MAX
After pushing softkey MIN / MAX, two peak detectors are operative for each
channel.
Glitches in the A channel as well as in the B channel can be catched at the same
moment,
since samples for both channels are taken at the same.moment.
When switching from 5 us/div to 2 us/div with MIN / MAX on the MIN / MAX
limitation is indicated by a message
No MIN / MAX possible when time base faster than
5 us/div. MIN / MAX switched off.
When switching from 2 us/div to 5 us/div and MIN / MAX on, the following
message is displayed
Previous MIN / MAX setting restored.
- Glitches longer than 3 ns are detected with an accuracy of more than 50 %.
- Glitches shorter than 20 ns are not detected if occurring during the reset time of
the peak detectors.
-
MIN
/ MAX
is not possible in combination
with the AVERAGE
function,
If MIN / MAX is selected when AVERAGE
is active,
a message
MIN / MAX and AVERAGE
not possible.
MIN / MAX on, AVERAGE off.
is displayed.
7
2
BW
LIMIT
After pushing softkey BW LIMIT a bandwidth filter is active. This is indicated by an
intensified text BW LIMIT.
The bandwidth is 3 dB down at20 MHz, using a 6 dB/octave filter. The filter is active
in both channels.
7
3
AVERG
OFF
After pushing softkey AVERG
OFF, AVERAGE
is switched
off.
This function
is only displayed when AVERAGE
is selected.