4-20E
l¥ 50 Ohmis on and an input voltage higher than 25 V dc andacr-m.s. is applied to the input
socket; 50 Ohm
input impedance will not be switched off and the message
OVERLOAD
ON CHANNEL A: reduce signal to prevent damages.
FER
III
ISI TI II
III IRI IE II I III II II REI
RT IR.
is displayed.
IMPORTANT:
Remove the input voltage or decrease the input voltage immediately
to <5 V de and ac r.m.s.!
4
OFFSET>
(channel A)
!f OFFSET is selected the VERTICAL/OFFSET A menu
is displayed
and an offset (vertical shift before attenuator) value can be selected.
| OFFSET
is only effective if DC input coupling is selected.
VERTICAL
it OFFSET
is selected and AC coupling is active, a message
OFFSET A
AUTO
4
OFFSET only possible if channel is DC- coupled.
2
AC is switched to DC.
ZERO
Cs
;|
is displayed and the input coupling is automatically switched to DC.
ENTER
>
4
:
+
omy
pom
uP
52}
DOWN
s
SCREEN DIV
7
RETURN
8
4
1
AUTO
With softkey AUTO, the offset value is set to such a level, that the mid—value of the
input signal is shified as much as possible to mid-memory.
A message
* * * AUTO
OFFSET
FOR CHANNEL A * * *
ee
es
is displayed.
The new offset value is displayed in the softkey text area. The VERTICAL SHIFT is
set to zero.
A message
AUTO OFFSET error : signal offset out of range.
is displayed ii the input signal is exceeding the offset contro! range. In this case the
original offset value is restored.
4
2
=
4
3
ZERO
With softkey ZERO, the offset value is set to zero.