Time Base - Philips PM3320 Operating Manual

250 ms/s dual channel digital storage oscilloscope
Table of Contents

Advertisement

6.3.18
6.3.19
6.4
6.4.1
6.4.2
CHARACTERISTIC
Average
Constant is max.
Cross talk
(according to
IEC 351)
TIME BASE
Modes
Time
coefficients
-In recurrent
-In single scan
and multiple scan
-In single shot
and multiple shot
-In roll mode
-With
EXT CLOCK
-Error limit
(Ambient
15.35 °C)
In equivalent
time mode
In real time mode
SPECIFICATION
64x
32 x or off
<-30 dB @ 100 MHz
<-50 dB @ 2 MHz
Recurrent
Single shot
Single scan
Multiple shot
Multiple scan
Roll
5 ns/div..5 s/div
5 ns/div...100 ns/div
200 ns/div...5 s/div
50 ms/div...360 s/div
Depending on clock
frequency
6-9E
ADDITIONAL INFORMATION
Average formuia after the
first front change
(MIN / MAX switched off).
Sn-So(n-1)
So(n)=So(n-1)+ ——-————_
c
In ROLL mode.
Up to 4 shots,
Up to 4 scans.
Can be stopped manually or
by trigger.
Input via EXT CLOCK,
every
clock pulse a sample is
taken, so for single
channel 4k samples are
stored and for dual
channel
2 times 2k samples.
Add 2 % for ambient:
0..50°C.
Add 0,5 % for ambient:
0..50°C.