12.1. ESSENTIAL PERFORMANCE

The FT-245 should not ensure continuity of service. In the event of electrostatic discharge, the machine may reset
itself, interrupting treatment and returning to the homepage without delivering it and without losing the preset programs
and those saved by the user.
If the touchscreen freezes during the delivery phase, the user must press the 'Emergency stop' button, interrupting the
power supply (Figure 2, see 2.1). To rearm, the device must be restarted, passing all the initial checks and returning
to the homepage, without losing the preset programs and those saved by the user.
In the event of induced interference (provided these are within the levels for which the device was designed), the way
the device behaves may be similar to that described above for electrical discharge.