3. INSTALLATION

Please make sure that you have all the required parts
and tools to complete the installation of your device prior
to starting work.
Use of controls and settings and following
procedures other than those specified may result
in a risk of explosions and/or uncontrolled laser
radiation.

3.1. INSTALLATION AND CONNECTIONS

The FT-245 is delivered in a cardboard box complete
with internal packaging to protect the equipment during
transport and storage. Before removing the machine from
the box, place the container on a hard, flat surface, open
the top of the box, remove the top layers of protective
packaging and carefully take out the device and all of
its accessories. Inside the box, next to the laser are the
handpiece and all the accessories.
It is the customer's responsibility to keep the original
packaging, which must be used to protect the equipment
during transport, in the event it must be returned to EMS
for support, or for any other reason.
Once the device has been removed from the packaging
and placed on a hard, flat surface, proceed as follows:
1. Check that the handpiece is correctly positioned on
the handpiece stand.
2. Insert the ignition key provided into the power switch
located on the back of the device (Figure 3), keeping
it in position O. The ignition key can only be removed
in position O.
3. Check that the emergency stop button is not pushed
down. If in doubt, turn the red knob counterclockwise,
in the direction indicated by the arrows found on it.
In case of emergency, the emergency stop button
located on the front must be pushed, immediately
deactivating the laser device (Figure 2). To restart
the
device,
counterclockwise, in the direction indicated by the
arrows found on it.
10
00_FB-1007_16_rev_C_DOLORCLAST High Power Laser_EMS.indb 10
00_FB-1007_16_rev_C_DOLORCLAST High Power Laser_EMS.indb 10
simply
turn
the
red
4. Insert the remote interlock connector (see section 10.1)
into the corresponding jack located on the back of the
device (Figure 3).
5. Insert the female connector of the power cord into
the power socket located on the back of the device
(Figure 3).
6. Connect the plug of the power cable to an electrical
socket within the installation area.
To install the device on the cart (optional), please refer
to the assembly documentation attached to the latter.

3.2. REMOTE INTERLOCK

The remote interlock connector (Figure 3) is used to
interrupt the emission of the laser beam in case of certain
events external to the device.
Specifically, all access doors in the area where the device
is installed must be equipped with micro-interrupters with
normally open contacts (which close when the door is
fully closed), connected in series to pins 1 and 2 of the
corresponding connector; therefore, when one of the
doors is opened, an 'interlock alarm' situation arises
which results in the immediate interruption of any laser
emission in progress.
Do not apply a voltage to the remote interlock
contacts. The contacts of the remote interlock jack
must be closed to enable the system to operate.
The remote interlock connector is supplied with
pins 1 and 2 short-circuited to facilitate set up and
initial start-up. It is the customer's responsibility to
remove the jumper and connect them to the
protection devices of the room's entry points.
knob