Floor Fastening With Adapter Plate - Dentsply Sirona Sinius Installation Instructions Manual

Hide thumbs Also See for Sinius:
Table of Contents

Advertisement

4 Installation
4.2 Fastening of the base of the chair to the floor
20
5. For concrete floors only: Fasten the chair base with 5 screws (A) and
2 heavy-duty anchors (B). The heavy-duty anchors (B) must be
tightened with a torque of 20 Nm.
Removing transport lock on the footrest bracket
D
➢ Remove the cable tie (D) from the footrest bracket.
4.2.3

Floor fastening with adapter plate

Usage
The adapter plate should be used
● If the treatment center is being installed to replace an M1 that was
previously installed in the same location.
You can use the existing drill holes for fastening. Two additional drill
holes are required only in the area of the water unit.
● To compensate for uneven flooring.
Preparation
1. Line up the adapter plate with the corner valves and align the adapter
plate dimensionally as well.
2. Mark the fastening points.
3. Remove the adapter plate.
4. Drill the holes for the chair base and water unit (observe section
"Stability [ → 17]"!).
Installation Instructions Sinius / Sinius CS / Sinius TS
D3561.031.01.15.02 07.2018
Dentsply Sirona
63 22 627 D3561

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sinius csSinius ts