Troubleshooting
Error
Imaging software fails to rec-
ognize the unit
Error during data transmission
between unit and computer.
Error message "CRC error
timeout"
Software message: "VisionX
has detected that the phos-
phor storage plate may have
been exposed from the wrong
side. Please check the orien-
tation and the image quality
before making a diagnosis"
"The selected unit cannot cur-
rently be accessed. Please
check that the unit is switched
on and correctly connected."
13.3
Fault on the unit
Error
Unit does not switch on
42
|
EN-US
Possible cause
Unit not switched on
Connecting cable between unit
and computer not correctly con-
nected
Computer does not detect any
connection to the unit
Hardware error
The IP address of the unit is
being used by another unit
Connecting cable used is incor-
rect or too long
The phosphor storage plate was
exposed on the back (inactive)
side while the X-ray was being
taken
Connection to the unit inter-
rupted while the software was
still attempting to communicate
with the unit.
Possible cause
No mains voltage
On/off switch is defective
Remedy
Switch the device on.
❯
Check the connecting cable.
❯
Check the connecting cable.
❯
Check the network settings
❯
(IP address and subnet
mask).
Contact technician.
❯
Check the network settings
❯
(IP address and subnet mask)
and assign a unique IP
address to each unit.
Inform a service technician, if
❯
the problem persists.
Only use original cables.
❯
When diagnosing the X-ray
❯
image, note that the X-ray
image is displayed mirror-
inverted.
Restore the connection to the
❯
unit.
Repeat the process.
❯
Remedy
Check the mains cable and
❯
plug connection and replace if
necessary.
Check the power supply unit.
❯
If the green status LED does
❯
not light up, replace the
power supply unit.
Check the mains fuse in the
❯
building.
Contact technician.
❯