Select the Drivers tab.
❯
Click Driver Installation.
❯
Click OK to confirm.
❯
The Driver Setup installation wizard opens.
Follow the instructions of the installation wiz-
❯
ard.
Installing the calibration data
Every sensor requires the individual cali-
bration data stored on the enclosed data
carrier.
Insert or plug the enclosed data carrier in the
❯
drive.
The data carrier starts automatically and the
calibration data are installed.
If the data carrier fails to start automatically, run
❯
the setup.exe file.
2121100020L29 2101V007
8
Commissioning and first
start-up
NOTICE
Short circuit due to the build up of
condensation
Strong temperature fluctuations can
damage the unit.
Start-up the unit only once it has equi-
❯
librated to room temperature.
Do not expose the unit to strong tem-
❯
perature fluctuations (heating max.
3 °C/min). Exception: Standard heating
from room temperature approximately
20° C to body temperature approxi-
mately 37 °C during application.
Before starting every time, check the entire
❯
system for damage.
Perform and document safety checks in
❯
accordance with national regulations
(VDE 0751‑1, IEC 60601‑1).
Carry out and document an acceptance test in
❯
accordance with national regulations.
The unit supports the following imaging pro-
grams:
– VisionX manufactured for Air Techniques
– DBSWIN manufactured for Air Techniques
– Please inquire about third-party software
Always use the current version of the
imaging program in the commissioning of
the device. Check the version of the
enclosed imaging program versus the ver-
sions available at www.airtechni-
ques.com.
8.1
Configuring the unit in
VisionX
The configuration is done in the settings of the
imaging software.
For further information, see the Help func-
tion of VisionX (E7303).
Requirements:
ü SensorX is connected to computer.
Start-up SensorX.
❯
Click on
once.
❯
Click Devices.
❯
Click on the SensorX unit in the list.
❯
Installation
EN-
US
19