8
Commissioning and first
start-up
NOTICE
Short circuit due to the build-up of
condensation
Do not switch on the unit until it has
❯
warmed up to room temperature and
is dry.
The unit supports the following imaging pro-
grams:
– VisionX made by Air Techniques
– VisionX Connect
– Third-party software on request
Always use the latest version of the
imaging program and ScanX ServiceTool
when commissioning the device.
8.1
Configuring the network
Network configuration
Various options are available for network configu-
ration:
ü Automatic configuration via DHCP.
ü Automatic configuration via Auto-IP for direct
connection of device and computer.
ü Manual configuration.
Configure the network settings of the device
❯
using the software or, if applicable, the touch
screen.
Check the firewall and release the ports, if
❯
applicable.
Network protocols and ports
Port
Purpose
1900 UDP
Device identification
80 TCP
Device identification
438 TCP
Unit data
22 TCP
Diagnosis
n/a
Check that the device
is switched on
When the unit is first connected to a com-
puter, it applies the language and time
settings of the computer.
2144100140L29
G8802
Service
SSH
ICMP /
ping
2310V007
8.2
Configuring the unit
Use the ScanX ServiceTool for configuration.
To start the ServiceTool via VisionX :
❯
Select
> Devices > Configuration > Serv-
ice > Service Tool.
Alternatively: Launch the ServiceTool via the
Windows Start menu:
Start > VisionX ServiceTool > VisionX Serv-
iceTool
Mark the connected unit in the list.
❯
If the connected device is not included in
the list, check if the device is switched on
and connected to the network. Then click
Find again.
Click OK.
❯
If connection fails, an error is displayed.
Select workflow 001.
❯
Follow the instructions provided by the Service-
❯
Tool.
Entering a permanent IP address (recom-
mended)
To reset the network settings, keep the
unit reset key pressed for 15 - 20 sec-
onds while switching on.
Select Network settings.
❯
Change Use DHCP to off.
❯
Enter the IP address, subnet mask and gate-
❯
way.
Click Save changes.
❯
The configuration is saved.
8.3
Security settings
Communication between imaging software and
device is always encrypted. At the time of deliv-
ery, communication is protected by a default
password: 123456. If the security requirements
are more extensive, this password needs to be
changed in the settings of the imaging software.
For more information, please refer to the manual
of the imaging software.
8.4
Testing the device
You can scan an X-ray image to check if the unit
is properly connected.
Open VisionX.
❯
Create an X-ray station for the connected unit.
❯
Log-in the demo patient (patient ID:
❯
DEMO0001).
Installation
EN-US
|
23