Usage
Store the dust cover in a clean place when it is
❯
not in use.
34
|
EN-US
11 Cleaning and disinfection
When cleaning and disinfecting the unit and its
accessories, comply with national directives,
standards and specifications for medical prod-
ucts as well as the specific specifications for
dental practices or clinics.
NOTICE
The use of unsuitable agents and
methods can damage the unit and
accessories
Only use the disinfection and cleaning
❯
agents specified or approved by Air
Techniques and the EPA.
Comply with the operating instructions
❯
of the disinfectants and cleaning
agents.
Wear hand protection.
11.1
Phophor storage plate scan-
ner
Surface of the unit
The plate guide needs to be removed
prior to cleaning and disinfection (see
"10.2 Replacing the plate guide").
For cleaning and disinfection of the plate
guide, see "Plate guide".
The surface of the unit must be cleaned and dis-
infected if it is contaminated or soiled.
NOTICE
Liquid can cause damage to the unit
Do not spray the unit with cleaning
❯
agents or disinfectants.
Make sure that liquid penetrates into
❯
the unit.
Remove any soiling with a soft, lint-free cloth
❯
that has been dampened with cold tap water.
Disinfect the surfaces with a disinfectant wipe.
❯
Alternatively, use a spray disinfectant on a soft,
lint-free cloth. Comply with the operating
instructions of the disinfectant.
2144100140L29
G8802
2310V007