Scanning The Image Data Via A Computer - Air Techniques ScanX Swift 2.0 Installation And Operating Instructions Manual

Table of Contents

Advertisement

Usage
10.4
Scanning the image data via
a computer
Starting the phosphor storage plate scanner
and software
The reading-out process is described
using the VisionX imaging software.
For further information regarding the use
of the imaging software, refer to the rele-
vant manual.
Start VisionX.
Select the patient.
Select the corresponding image type in the
menu bar.
Select the unit.
Set the acquisition mode.
Recording starts immediately.
If ScanManager is enabled, select the X-ray job
on the touch screen of the unit.
Result:
The touch screen will display an animated visual
symbol requesting insertion of the phosphor stor-
age plate.
Insert the phosphor storage plate only
once the bar above the animation is
green.
Do not insert another phosphor storage
plate as long as the bar of the animation
is blue.
Scanning the PSP
Observe the information on the display.
When feeding in the phosphor storage
plate, make sure that it is correctly
assigned to the right patient.
Tear the barrier envelope lengthwise at the
notch.
32
|
EN-US
Place the barrier envelope containing the phos-
phor storage plate in the plate guide such that
it is aligned in the middle and straight. The
opened side of the barrier envelope faces
down, the inactive side of the PSP faces the
operator.
The device recognizes automatically if the
phosphor storage plate was fed in wrong
orientation (active side toward the user)
and displays a message to this effect on
the touchscreen. Turn over the phosphor
storage plate (in active site toward the
user)and feed it in again immediately.
The phosphor storage plate must not be
pushed out of the barrier envelope before plac-
ing it against the plate guide. Otherwise there is
a risk that image information could be erased
by ambient light (see "9 Correct use of phos-
phor storage plates").
Slide the phosphor storage plate out of its bar-
rier envelope downwards into the device until
the phosphor storage plate is automatically
drawn in.
The barrier envelope is held in place by the
plate guide and is not transported into the unit.
Make sure to insert only the phosphor storage
plate, and not the barrier envelope, into the
unit.
The image data is automatically transmitted to
the imaging software.
After it has been scanned, the phosphor stor-
age plate is erased and drops into the collec-
tion tray.
Remove the empty barrier envelope.
2144100140L29
G8802
2310V007