10.1

Switch the device on

Tapping the On/Off switch
device.
The On/Off button lights up briefly and the
device starts up.
As soon as the device is ready for use the
On/Off button is blue and the start screen is
displayed.
10.2

Replacing the plate guide

The device can read phosphor storage plates in
the sizes S0 to S4. The matching plate guide is
needed for each size of phosphor storage plate.
The size of the phosphor storage plate is marked
on the plate guide.
CAUTION
Loss of image information and equip-
ment damage if the wrong plate guide
is used
Always use the plate guide of the cor-
rect size for the PSP being used.
Before each scanning process, com-
pare the phosphor storage plate size
with the markings on the plate guide.
2144100140L29
G8802
turns on the
2310V007
The plate guy can be switched at any
time. To prevent a loss of image quality
do not switch the plate guide during a
scanning process.
Use your finger to press the indentation while
tilting the plate guide toward the front at the
same time.
Insert the plate guide from above.