Installation
Only qualified specialists or persons
trained by Air Techniques may install,
connect, and commission the unit.
6
Requirements
6.1
Installation/setup room
The room chosen for set up must meet the fol-
lowing requirements:
– Closed, dry, well-ventilated room
– It should not be a room made for another pur-
pose (e.g. boiler room or wet cell)
– Max. light intensity 1000 Lux, no direct sun-
light at the place of installation of the unit
– There should be no major fields of interference
(e.g. strong magnetic fields) present that can
interfere with the proper operation of the unit.
– Ambient conditions correspond to "4 Techni-
cal data".
6.2
System requirements
For the system requirements of the com-
puter systems, visit the download area at
www.airtechniques.com (document no.
E7201).
6.3
Monitor
The monitor must meet the requirements for digi-
tal X-ray with a high light intensity and wide con-
trast range.
Strong ambient light, sunlight impinging directly
onto the monitor and reflections can make it
more difficult or even impossible to perform a
diagnosis based on the X-ray images.
2144100140L29
G8802
2310V007
7
Installation
7.1
Setting up the unit
NOTICE
Damage to sensitive components of
the unit due to shocks or vibrations
Do not expose the unit to any strong
❯
vibrations or shocks.
Do not move the unit during operation.
❯
Portable and mobile HF communication applian-
ces can interfere with the effectiveness of electri-
cal medical devices.
Do not stack the unit next to or together with
❯
other appliances.
If, however, this unit is operated next to other
❯
units or stacked with other units, monitor the
unit carefully in the configuration selected in
order to ensure normal operation.
The unit can be set up as a tabletop unit or
mounted on a wall using the wall bracket.
The load-bearing capacity of the table or wall
must be suitable for the weight of the unit (see
"4 Technical data").
Setting the unit on a table
To prevent errors when scanning the
image data, install the unit so it is not
exposed to vibrations.
Place the unit on a firm, horizontal surface.
❯
Installing the unit with the wall mounting
bracket
The unit can be mounted on a wall with the wall
mounting bracket (see "3.3 Optional items").
Installation
5,1 kg
11,24 lb
EN-US
|
19