22. ERROR MESSAGES
I
Display
1
Meaning
1
Action
I
Not formatted
New card. Card has not
Format the card.
been formated.
1
Out of range
Out of range
X >
i
<Y
During REM, the verti-
cal a n g l e is m o r e
thani89" or the mea-
sured distance
i s
more
than 9999.999m
Tilt sensor range error.
Tilt angle exceeds i 3 ' .
Press
t o s t o p
measuring.
Level the SET2C again.
Record error
External
device
does
n o t reply w i t h ACKI
NAK.
(when "recording" pa-
rameter is set to "out".
Check to see if there are
any abnormalities i n
cables
o r
e x t e r n a l
equipment, or if there is
a p r o b l e m w i t h t h e
program.
Signal off
Tilt error
A t start of measure-
ment, the returned sig-
nal was totally absent
or disturbed.
While setting the azi-
muth angle, tilt angle
exceeds rt3'.
Time out
Sight the target again.
Measure again after
c o n f i r m i n g
t h e
r e -
turned signal using the
signal checking mode.
Level the SET2C again.
Tilt
Out of range
During distance mea-
surement, tilt angle ex-
ceeds i 3 ' .
No measured distance
dataisreceivedwithin2
minutes of starting the
measurements, or the
m e a s u r e d
distance
d a t a c a n n o t b e o b -
tained for a total of one
minute.
Level the SET2C again.
Sight the target again.
Measure again after
c o n f i r m i n g t h e
re-
turned signal using the
signal checking mode.