CHAPTER
16
Configuring the system for automatic bump testing
If the system or one of its cradles is not configured for automatic bump testing, you must
configure it. This enables automatic bump testing for a stand-alone cradle or any number of
cradles in a system.
1. Connect a PC running Safety Suite Device Configurator (SSDC) to the AutoRAE 2 Cradle via
a USB cable.
2. Press and hold the Bump key until "PC" appears in the display.
3. Start Safety Suite Device Configurator (SSDC).
4. Login with your username and password. By default the user name is "administrator" and
the password will be "Default123".
5. Click "LOG IN"
6. Select the AutoRAE 2 Cradle.
7. Select "Automatic Bump Test."
AutoRAE 2
AUTOMATIC BUMP TESTING
62
User Manual