Connecting calibration gas to a Cradle

The AutoRAE 2 Cradle can accommodate two gas calibration gas cylinders (mixture or single
gas in each). In addition, there is a connection labeled "Exhaust," for venting the gas after it has
gone through the AutoRAE 2 Cradle. All gas connections are barbed to secure the hoses to
them. Appropriately non-reactive/non-adsorptive tubing with a 1/8" I.D. should be used (Teflon
for PID or corrosive or reactive gases, Tygon for others). The cylinders must have demand-flow
regulators (0 to 1,000 psig/70 bar) installed.
No.
1
2
3
IMPORTANT!
Always check that the active gas configuration on the AutoRAE 2 Cradle and the
type/concentration of the actual calibration gases connected to the Cradle match before you
begin any bump test or calibration.