Bump Testing - Honeywell AutoRAE 2 Quick Start Manual

Hide thumbs Also See for AutoRAE 2:

Advertisement

by Honeywell

Bump Testing

You can bump test all or individually selected instruments that are seated in the system's cradles.
From the main screen, enter Bump Test by pressing [Y/+] twice ("Function" and then "Select").
Note: Any cradle that does not have an instrument in it is shown with a gray checkbox. You cannot check/uncheck
this box.
To select all instruments in cradles to bump test:
1. Press [Y/+]. When you check "Bump All," the "Bump All" checkbox and all of the other checkboxes for
cradled instruments are automatically checked.
2. Press [MODE] for "Done."
3. Press [Y/+] to start.
To select specific instruments to bump test:
1. Press [N/-] to scroll down the list to an instrument you want to select for bump testing.
2. Press [Y/+] to toggle the selection between unchecked and checked.
3. Press [MODE] for "Done."
4. Press [N/-] to select "Exit," and then press [Y/+] to exit.
If you did not press [Y/+] to start the process, then a countdown begins. When the countdown reaches zero, then
the instruments are bump tested.
To abort a bump test: Press [N/-]. You will be asked if you want to abort the process. Press [Y/+] for "Yes" and
[N/-] for "No."
IMPORTANT!
If an instrument does not pass a bump test, a full calibration is automatically started.