Gas Inlet Configuration Settings - Honeywell AutoRAE 2 User Manual

Hide thumbs Also See for AutoRAE 2:
Table of Contents

Advertisement

If you want to synchronize the date and time on the AutoRAE 2 Cradle with the time on the PC,
click the box labeled "Synchronize with Server."

Gas inlet configuration settings

Gas Inlet windows tells the AutoRAE 2 what kind of gas is supplied to each gas inlet.
The Gas Inlet windows section includes configuration parameter settings for the two AutoRAE 2
Cradle gas inlets including gas types, concentrations, concentration units, purge time, and
soak time for gas cylinders connected to each gas inlet. You can modify these values and
upload them to your AutoRAE 2 Cradle(s) or download the values currently programmed into
the Cradle to Safety Suite Device Configurator (SSDC)
Gas Inlet windows cover settings for two cylinders of gas corresponding to gas inlets 1 and 2,
respectively, on the side of the AutoRAE 2 Cradle.
The window shows the Gas Inlet number, the gasses included, Gas Lot Number, Gas Expiration
Date and data that includes Gas Name, Concentration, Concentration Unit, Purge Time (Sec.),
and Soak Time (Sec.).
AutoRAE 2
54