Troubleshooting
Fault description
N > ROTOR MAX
The left half of
the display lights up.
9.2

Performing a MAINS RESET

9.3

Emergency release

36 / 39
Cause
The rotor has been changed.
The built-in rotor has a higher
maximum speed than the pre-
viously used rotor, and it has
not yet been detected by the
rotor detection function.
-
1.
Set the mains switch to
2.
Wait 10 seconds.
3.
Set the mains switch to
The lid cannot be unlocked by the motor in the event of a power failure.
Emergency unlocking by hand must be performed.
WARNING
Risk of electric shock due to maintenance and servicing
work on live device.
Disconnect the device from the mains before carrying
out repairs and maintenance.
WARNING
Danger of cutting and crushing due to moving rotor.
Do not open the lid until the rotor has stopped.
Personnel:
Trained user
Fig. 17: Emergency release
1
Hole
1.
Look through the window in the lid to ensure that the rotor is sta-
tionary.
Rev.: 09 / 11.2023
Remedy
Set a speed up to the maximum
speed of the previously used rotor.
[START/IMPULS] button to
Press the
perform rotor detection.
Notify customer service.
[0] .
[I] .
1
AB4600en