Message displays
The following table lists LCD messages and associated solutions.
Display
INFO:001
Panel load. OK?
INFO:002
Panel loading...
INFO:003
Enter panel name.
ESC: CANCEL,
ENTER: SAVE EXEC
INFO:004
Enter panel name.
Panel is used, will be overwritten.
ESC: CANCEL,
ENTER: SAVE EXEC
INFO:005
Panel saving...
INFO:006
Clear panel. OK?
INFO:007
Panel clearing...
INFO:008
Printing...
INFO:010
Start interval print.
INFO:011
Stop interval print.
INFO:020
Performing 0 adjustment. OK?
INFO:021
Clear 0 adjustment data. OK?
INFO:022
Cleared 0 adjustment data.
INFO:023
0 ADJ warning.
Adjust within 1% f.s.
INFO:025
Undo statistical calculations.
INFO:026
Self-calibrating...
INFO:030
Reset?
NORMAL RESET (without panel clear)
/ SYSTEM RESET (with panel clear)
/ MUX RESET (only CH settings)
INFO:035
MUX CH settings will be reset. Change
setting?
INFO:036
0 adjusting...
INFO:037
Short-circuit pin No.1 to No.42, OK?
INFO:038
Testing MUX units...
INFO:040
Enter password for Adjustment Mode.
INFO:041
Password is wrong.
INFO:080
Self-calibration is set to "manual".
Panel data will be loaded. Continue?
Panel data is being loaded.
Enter a name for the panel being saved.
Cancel the save operation with the
with the
ENTER
Enter a name for the panel being saved.
The specified name already exists and will be overwritten if you
proceed.
Cancel the save operation with the
with the
ENTER
Panel data is being saved.
Panel data will be cleared. Continue?
Panel data is being cleared.
Printing in progress.
Interval printing started.
Interval printing stopped.
Zero adjustment will be performed. Continue?
Zero adjustment values will be cleared. Continue?
Zero adjustment data was cleared.
Zero adjustment data values are large. (Warning)
→ It is recommended that values be within 1% of range full-scale.
One statistical calculation was undone.
Self-calibration measurement in progress.
The instrument will be initialized.
The MUX channel settings will be initialized when switching
between 4-terminal and 2-terminal measurement.
Zero adjustment is being performed with MUX scanning.
To perform the unit test, short all the pins numbered 1 to 42.
The multiplexer unit test is being performed.
Enter the password for adjustment mode.
The adjustment mode password is incorrect. Please enter the
correct password.
Self-calibration measurement is set to MANU.
HIOKI RM3545E965-00
22
Description
ESC
key or save the panel
key.
ESC
key or save the panel
key.