Printer issues
Issue
No data is being printed. The instrument and printer are not
Printed text is garbled.
Multiplexer issues
Issue
It is not possible to
switch to the multiplexer
inputs.
Channels cannot be
switched by operating
the instrument's keys.
Channels cannot be
switched with EXT. I/O.
Measured values are
unstable.
The measured value
differs from the expected
resistance value.
Possible causes
connected properly.
The printer and instrument settings do not
match.
Display
Possible causes
[ERR:60]
Measurement leads
are connected to the
measurement terminals on
the front of the instrument.
[CH]
is
The front terminals
not being
are being used as the
displayed.
measurement terminals.
Scan display
The scan function is set to
(list display)
auto or step.
The set unit number and
the unit number in which
the Z3003 is installed differ.
[RMT]
The instrument is
in remote mode, in
which it is controlled
by communications
functionality.
The MUX signal is not on.
See "Measured values are unstable." (p.  1 2)
The wrong channel is being
measured.
The wiring is shorted.
The route resistance is too
high.
Measurement leads
are connected to the
measurement terminals on
the front of the instrument.
HIOKI RM3545E965-00
• Check whether the connector has been
connected.
• Check whether the interface settings have been
configured properly.
• If using the PRINT signal, see "Unable to print."
(p.  1 5)
Check the printer settings again.
Do not connect any measurement leads to
the measurement terminals on the front of the
instrument.
If
[ERR:60]
is displayed even though no
measurement leads are connected, turn off the
instrument and remove the Z3003. If
not displayed after removing the Z3003, the Z3003
could be broken. Request repair.
Set the measurement terminals to MUX.
Set the scan function to off in order to switch
channels with key operation.
Check the settings and the unit on the back of the
instrument.
Operate the instrument after canceling remote
mode.
Turn on the MUX signal.
Check the current channel and the channel setting.
Exercise care to avoid shorted wires.
For 2-wire connection, the route resistance
affects the measured value directly. Perform zero
adjustment.
Do not connect measurement leads to the
measurement terminals on the front of the
instrument when using the multiplexer.
18
Solutions
Solutions
[ERR:60]
is

Advertisement

loading