Display
ERR:091
RAM error.
ERR:092
Memory access failed.
Main power off, restart
after 10s.
ERR:093
Memory read/write
error.
ERR:095
Adjustment data error.
ERR:096
Backup data error.
ERR:097
Power line detection
error.
Select power line cycle.
ERR:098
Blown FUSE or
measurement lead is
broken.
ERR:099
Clock is not set.
Reset?(13-01-01
00:00:00)
Press F2"
ERR:100
MUX unit error.
Description
CPU RAM error
A communications error occurred
while attempting to access the
memory.
Memory read/write test error
Adjustment data error
Settings backup error
Power frequency detection error
The fuse has been tripped.
The clock is not set, so pressing F2
[OK]
displays the initialized time 13-
01-01 00: 00: 00.
The MUX unit experienced an error.
HIOKI RM3545E965-00
21
Remedy
The instrument is malfunctioning.
Request repair.
Turn off the main power switch, wait at least
10 seconds, and turn it back on.
The instrument is malfunctioning.
Request repair.
The instrument is malfunctioning.
Request repair.
Settings were reinitialized. Reconfigure
measurement conditions and other settings.
Set the frequency to match that of the power
being supplied to the instrument.
Replace the fuse.
If the fuse is not blown out, the measurement
and guard terminals may be shorted.
Disconnect the measurement leads and check
whether or not the error occurs. If the error
still occurs, request repair of the instrument.
Furthermore, be sure to use a Hioki-specified
fuse for replacement.
Specified fuse: F1.6AH/250 V (non arcing)
×
20 mm
5 mm dia.
The backup battery needs to be replaced.
Contact your authorized Hioki distributor or
reseller.
The instrument is malfunctioning.
Request repair of the instrument.