Error Handling And Diagnostics - Beckhoff EL3681 Documentation

Digital multimeter terminal
Table of Contents

Advertisement

Error handling and diagnostics

6
Error handling and diagnostics
Error **
Under-
Over-
Index
ange **
range **
Index
Index
0x6000:0
0x6000:01
0x6000:02
7 [} 158]
[} 158]
[} 158]
1
1
1
1
1
1
1
**) Error LED is lit, see LED description [} 48]
168
Data in-
TxPDO
Range in-
valid
State **
valid
Index
Index
Index
0x6000:04
0x6000:0
0x6000:05
[} 158]
F [} 158]
[} 158]
1
1
1
1
1
Version: 3.0
Error descrip-
Remedy
tion
Measurement is
If possible use a
below range
smaller measuring
range; otherwise
increase the input
value
Measuring
If possible use a
range exceeded
larger measuring
range; otherwise
reduce the input
value
Conversion
automatic (ADC
error
does not respond,
automatic restart)
Measuring
Select valid
range invalid
measuring range;
for Autorange
perhaps select
option "scaled" or
activate SAI Range
(outputs).
Data invalid
Wait (read ADC
Zero, startup or
conversion error)
Conversion
Switch on again
error after
power-up
EL3681

Advertisement

Table of Contents
loading

Table of Contents