Promax OD-603 Manual page 84

Digital oscilloscope
Table of Contents

Advertisement

Figure 78. Set the network parameters of the Oscilloscope Software.
Set the network parameters of the oscilloscope.
In
the
oscilloscope, press the Utility button and press H1 button, turn the M
knob to select the LAN Set; press the H2 button, the set menu is
displayed on the right. Set the IP and the Port to the same value as
the "Ports-settings" in the software in step (3). The Netgate and Net
mask should be set according to the router. Press the H3 button to
select "Save set", it prompts "reset to update the config". After
resetting the oscilloscope, if you can get data normally in the
oscilloscope software, the connection is successful.
Figure 79. Set the network parameters of the oscilloscope.
79
June 2016
Page