Promax OD-603 Manual page 82

Digital oscilloscope
Table of Contents

Advertisement

Figure 75. Set the network parameters of the Oscilloscope Software.
Set the network parameters of the oscilloscope.
In
the
oscilloscope, press the Utility button and press H1 button, turn the M
knob to select the LAN Set; press the H2 button, the set menu is
displayed on the right. Set the IP and the Port to the same value as
the "Ports-settings" in the software in step (3). Press the H3 button to
select "Save set", it prompts "reset to update the config". After
resetting the oscilloscope, if you can get data normally in the
oscilloscope software, the connection is successful.
Figure 76. Set the network parameters of the oscilloscope.
77
June 2016
Page

Advertisement

Table of Contents
loading

This manual is also suitable for:

Od-606

Table of Contents