Promax OD-603 Manual page 64

Digital oscilloscope
Table of Contents

Advertisement

The description of Pass/Fail Menu is shown as the follows:
Function Menu
operate
Output
Rule
SaveRule
Pass/Fail test:
Detect whether the input signal is within the limits of the rule, if it
exceeds limits of the rule, it is "Fail"; otherwise it is "Pass". Also it can
output fail or pass signal by built-in and configurable output port. To
run the test, read the following steps:
Press Utility button, then H1 button, turn the M knob to choose
Pass/fail menu option, Pass/Fail menu will be displayed in the bottom.
Enable switch on: Press H2 button to show Operate menu, then,
press F1 button to set Enable as ON.
Create rule: Press H4 button to enter Rule setting menu. Press F1
button to choose the source; Press F2 button, turn the M knob to set
Horizontal tolerance; Press F3 button, turn the M knob to set Vertical
tolerance; Press F4 button to create the rule.
Set output type: Press H3 button to enter Output option setting.
Choose any one or two of the options "Pass", "Fail" or "Beep". "Pass"
and "Fail" are mutually exclusive options, which could not be chosen
simultaneously. "Stop" means stop once the condition satisfies your
setting.
June 2016
Setting
Enable
Control enable switch.
Operate
Control operate switch.
Pass
Signal tested corresponds with the rule.
Fail
Signal tested not correspond with the rule.
Beep
Beep when it satisfies the rule.
Stop
Stop once satisfying the rule.
Info
Control the display status of info frame.
Source
Select source CH1, CH2 or Math.
Horizontal
Change the Horizontal tolerance value by turning
the M knob.
Vertical
Change the Vertical tolerance value by turning the
M knob.
Create
Use the rule set as testing rule.
Choose any one from Rule1~Rule8 as your rule
Number
name.
Save
Click Save to save the rule.
Load
Load some rule as the testing rule.
Description
59
Page