4.2.2

Connect through a router

Connection. Use a LAN line to connect the oscilloscope with a router,
the LAN port of the oscilloscope is in the right side panel; the
computer should be connected to the router too.
Set the network parameters of the computer. Since the
oscilloscope can not support obtaining an IP address automatically,
you should assign a static IP address. The Default gateway and Subnet
mask should be set according to the router. Here we set the IP address
to 192.168.1.71, Subnet mask is 255.255.255.0, Default gateway is
192.168.1.1.
Figure 77. Set the network parameters of the computer.
Set the network parameters of the Oscilloscope Software. Run
the software on the computer; choose the "Ports-settings" of the
"Communications" menu item. Set "Connect using" to LAN. About the
IP, the first three bytes is same as the IP in the step (2), the last byte
should be different. Here, we set it to 192.168.1.72. The range of the
port value is 0~4000, but the port which under 2000 is always be
used, so it is suggested to set it to the value above 2000. Here, we set
it to 3000.