RTLF / RKLF scale applicator

Suitable for:
RTLF tape scale, any length
RKLF narrow tape scale, any length
Required parts:
Appropriate length of scale (see
'RTLF installation drawing'
or
'RKLF installation drawing'
on page 12)
End clamp kit, RKLF only (A-9523-4015)
RGG-2 two part epoxy adhesive, RKLF only (A-9531-0342)
Datum clamp, RTLF only (A-9585-0028)
Loctite 435, RTLF only (P-AD03-0012)
Appropriate cleaning solvents (see
'Storage and handling'
Readhead mounting bracket (see table below)
Appropriate scale applicator (see table below)
2 × applicator mounting screws
2 × M2 × 6 screws, top mount applicator only
Optional parts:
End cover kit, RTLF only (A-9585-0035)
Renishaw scale wipes (A-9523-4040)
Lint-free cloth
Reference mark deselection stickers (A-9402-0049 for a sheet of 20)
Loctite 435 dispensing tip (P-TL50-0209)
16
ATOM linear encoder system
Readhead mounting bracket and scale applicator:
There are different versions of the applicator depending upon the scale and design of the readhead mounting bracket.
For more information on designing your bracket, contact your local Renishaw representative.
on page
11
Mounting
on page 8)
Applicable
bracket
RTLF
RKLF
Slim applicator
Side mount applicator
In place of the
In place of the readhead bracket
readhead bracket
Side mounting bracket
'L' mounting bracket
(A-9401-0103)
(A-9402-0037)
Customer-designed bracket
Customer-designed bracket
A-9402-0060
A-9402-0034
A-6547-1947
A-6547-1943
Top mount applicator
In place of the readhead
'L' mounting bracket
(A-9402-0037)
Customer-designed bracket
A-9402-0028
A-6547-1939