Endura ACA592
Toroidal conductivity transmitter
7.3.3 Device Setup
Menu
Device Setup
Exit
Device Setup Level
Parameter
Main Sensor
Analyzer Type
Conc. Soln.
Conc. Unit.
Conc. Curve Name
Conc. Curve Table
OI/ACA592/TC–EN Rev. B
Used to set the sensor / sensor input, run temperature sensor
3
/ sensor recognition and obtain transmitter information.
Note. Required level of access = Advanced.
Select
Description
Sets the analyzer to a measurement type:
Conductivity
Concentration
Enabled if Analyzer Type = Concentration.
Sets the concentration solution:
0–18% HCI, 0–20% H
SO
2
Enabled if Analyzer Type = Concentration.
Selects the concentration unit:
%, ppm, µg/l, ppb, mg/l or none
Enabled if Analyzer Type = Concentration.
A user-defined name.
Press the
key to enter the required characters for the name as detailed on page 40.
Enabled if Analyzer Type = Concentration.
Enables entry of 6-point conductivity / concentration curve for user-defined concentration
calculation.
Range = X1 to X6 conductivity in uS/cm, Y1 to Y6 concentration
, User Def., 0–1% NaOH or 0–20% NaC1.
4
7 Operator Pages and Menus