7.3 ACA592 Toroidal Conductivity Transmitter Parameters

7.3.1 Easy Setup

Menu
Easy Setup
Exit
Easy Setup Level
Parameter
Analyzer Type
Conc. Soln.
Conc. Unit
Conc. Curve Name
Temp. unit
Temp. Comp. Type
Manual
Auto
52
Users with Advanced level access have read and write
1
privileges to the parameters listed.
Users with Standard and Read Only access have read-only
privileges to parameters at Easy Setup level.
Select
Description
Selects the measurement type:
Conductivity
Concentration
Enabled if Analyzer Type = Concentration.
Sets the concentration solution from the following options:
0–18% HCI, 0–20% H
SO
2
Enabled if Analyzer Type = Concentration.
Selects the concentration unit:
%, ppm, µg/l, ppb, mg/l, none
Enabled if Analyzer Type = Concentration.
A user-defined name.
Press the
key to enter the required characters for the name as detailed on page 40.
(Temperature Units)
Sets the temperature unit (°C or °F) applied to all transmitter temperature parameters and
values.
(Temperature Compensation Type)
Manual temperature compensation is enabled when the device does not detect any
temperature sensor connection either at power-up or upon execution of Run TC
Recognition. When Temp Comp Type is Manual, the user configurable Manual Temp value is
used for temperature compensation.
Enabled when a temperature sensor is detected by the device either at power-up or when
executing the Run TC Recognition (see page 56) routine.
, User Def., 0–15% NaOH or 0–20% NaC1.