Toroidal conductivity transmitter
Device Setup Level
Parameter
Device Info
Device Type
Device Serial No.
Software Version
Hardware Version
HART Tag
HART Descriptor
Long Tag
Write Protection
HW write prot.
SW write prot.
OI/ACA592/TC–EN Rev. B
Description
The type of transmitter, for example ACA592–TC (toroidal conductivity transmitter) –
read-only.
The device-specific serial number – read-only.
The transmitter's software version number – read-only.
The transmitter's hardware version number – read-only.
A configurable 8-character HART tag. Default tag is ACA592TC
Use to specify a short HART description for the transmitter (entered from the alphanumeric
character set).
Use to specify a longer HART description for the transmitter (entered from the alphanumeric
character set, maximum 32 characters).
Displays hardware / software write protection status.
Note. Configurations can be protected by enabling the configuration DIP-switch – see
Section 6.2, page 45.
Displays the current state of the Hardware Write Protection switch (Off / On).
Turns the software Write Protection Off / On.
Note. Software Write Protection cannot be modified when hardware write protection is
enabled through the DIP switch – see Section 6.2, page 45.