Installation
June 2023
3.5 
Secure the gas chromatograph (GC) to the
mounting bracket
Procedure
1. Screw two bolts, without the washers, into the top mounting holes on the back of
2. Maneuver the GC to insert the two top bolts into the eyelets of the mounting bracket
3. Screw in the two bottom bolts through the mounting bracket with the washers on.
4. One at a time, remove the top bolts, put on the washers, and screw the bolts into
36
the GC, leaving ½ in. (12 mm) of the thread exposed.
and allow the bolts to drop down and hold the GC loosely on the bracket.
The flat washer should be against the bracket, and the spring washer between the
flat washer and the bolt head. Tighten these two bolts by hand so that they secure
the GC in place.
the back of the GC and hand tighten.