15. Go to Tools → Cold Boot. Cold boot the GC.
16. Wait for the GC to boot.
17. Connect to the GC using MON2020.
18. Go to File → Restore Configuration (to GC) and restore configuration to the GC.
19. Wait for the heaters to stabilize.
20. In MON2020, go to Control → Auto Sequence to return the GC to normal operation.
Rosemount 470XA
The GC reboots automatically and disconnects from MON2020.
Troubleshooting
June 2023
133

Advertisement

Table of Contents