Using the Rosemount 470XA
June 2023
Figure 4-12: Valve Events Dialog Window Showing Cal Gas Save Events
By default, the Cal-Gas Save is switched on at 20 seconds and turned off at 200 seconds,
but you can change these values.
1. Start MON2020 and select Timed Events on the Application menu. The Timed Events
2. Locate the first Cal Gas Save event on the Valve Events table. This event switches on
3. Enter a new start time in the appropriate Time field.
4. Locate the second Cal Gas Save event on the Valve Events table. This event switches
5. Enter a new end time in the appropriate Time field.
6. Click OK to save the changes and close the window.
4.5.7 
New Module software assistant
When you install a new analytical module on the Rosemount 470XA and turn on the power,
the GC recognizes that a new module has been installed and starts the New Module
software assistant, which heats up the oven, runs carrier gas through the analytical paths,
and cycles the analytical valves to rapidly purge the system.
98
window opens.
the Cal-Gas Saver feature.
the Cal-Gas Saver feature off.