Emerson Rosemount 470XA Manual page 66

Gas chromatograph
Hide thumbs Also See for Rosemount 470XA:
Table of Contents

Advertisement

Using the Rosemount 470XA
June 2023
Procedure
1. Locate the switch at SW1 on the backplane. It is in front of the RJ-45 Ethernet plug.
2. Connect between the GC's RJ-45 connector and the local computer with a standard
3. Wait 30 seconds for the computer to update its TCP/IP settings.
4. Start Rosemount MON2020.
66
Flip the switch on.
This activates the GC's DHCP server. The server typically takes 20 seconds to initialize
and start up.
Figure 4-1: Backplane switches and RJ-45 connector
A. RJ-45 Ethernet plug
B. SW1 switch
NOTICE
If you intend to connect the RJ-45 port to a router or switch, ensure the GC's DHCP
server is turned off by flipping the SW1 switch to OFF. Connecting to your LAN with
the DHCP server on disrupts your local network's functioning.
NOTICE
You can connect the GC (or have it remain connected) to a local network via the
wired Ethernet 2 port on the back plane while the DHCP feature is being used.
Cat 5 Ethernet cable.
NOTICE
If your computer is not configured for dynamic IP addressing (static IP), you will
need to configure the GC with a static IP address that is in the same subnet as your
computer. See
Connect directly to a personal computer (PC) using the gas
chromatograph's (GC's) Ethernet
port.