Analog Output screens
Figure 4-8: Analog Output 1 screen
Figure 4-9: Analog Output 2 screen
Label
Switch
Variable
Current Value
Fixed Value
Zero Scale
Full Scale
mA
Rosemount 470XA
The name of the analog output as set in Rosemount MON2020.
An analog output has two operational modes:
Setting the switch to Variable means that the analog output is
proportional to the variable displayed in the Variables field.
Setting the switch to Fixed means that the analog output is set to the
value that is entered in the appropriate Fixed Value field.
Displays the system variable to which the analog output is associated. You
can change this variable with Rosemount MON2020.
Displays the current scaled value of the analog output signal.
If the analog output is set to Fixed, then analog output signal is set to the
value that you enter into this field. You can change this field.
The minimum analog output signal value.
The maximum analog output signal value.
Displays the amount of current being produced in milliamperes.
Using the Rosemount 470XA
June 2023
77