Using the Rosemount 470XA
June 2023
Carrier Pressure screen
Set Point
Current
Value
Status
Discrete Inputs screen
Label
Switch
74
Indicates the target pressure. This pressure is controlled by the external
regulator on the Rosemount 470XA unit.
Current pressure. The carrier pressure gauge value on the flow panel may
be slightly different from the value on the Carrier Pressure screen. Adjust
the carrier pressure regulator to the Set Point on this screen.
Indicates if the carrier pressure is within range of the set point.
If the screen displays
OK
If the carrier pressure is out of range, the screen displays an alarm message.
The name of the discrete input as set in Rosemount MON2020.
Indicates the discrete input's operational mode. There are three modes:
Auto: The gas chromatograph (GC) controls the discrete input's off or on
state.
Off
, the carrier pressure is within range.
Manual
MS-00809-0100-0470
Emerson.com