Troubleshooting
June 2023
6. On the new CPU board, set up switches as shown in the following image:
7. Install the new CPU board in the card cage.
8. Place the clear plastic cover back over the boards.
9. Close the GC cover.
10. Power up the GC and connect to it through MON2020.
11. In MON2020, go to Chromatograph → View/Set Date_Time. Set the date and time
130
Figure 8-1: CPU switch settings
A. Turn SW7 ON (toward the dot).
B. Turn SW3 OFF (away from the dot).
C. Turn SW4 OFF (away from the dot).
D. Turn SW6 OFF (away from the dot).
Ensure the board is seated firmly in place.
for the GC.