Advertisement

Installation
June 2023
NOTICE
You can establish up to ten simultaneous Modbus TCP connections from the Modbus
master. Connections attempts made after the tenth connection result in a No Response
error.
3.12.1 
Ethernet 1 port
Ethernet 1 was designed primarily for local connection to a computer, such as a
technician's laptop, for occasional maintenance and diagnostic purposes. The connector
is the same RJ-45 Ethernet connector commonly found on most Internet-capable devices.
NOTICE
If your computer is not configured to automatically configure Ethernet settings, contact
your IT department for instructions on how to change your IP settings to an address in the
same range as the Ethernet subnet on the gas chromatograph (GC) or to obtain an IP
address and subnet for the GC that will work with your computer's settings.
If wiring Ethernet 1 to other Ethernet-enabled devices, such as a router, hub, or local
area network, then set the DHCP server switch to OFF to ensure that the operation of the
network is not affected.
Related information
Connect directly to a personal computer (PC) using the gas chromatograph's (GC's)
Ethernet port
3.12.2 
Ethernet 2 port
Figure 3-15: Ethernet 2 port on the backplane
A. Ethernet port 2
B. Ethernet device
The second Ethernet port is intended to be connected to an Ethernet-enabled supervisory
network such as a flow computer, Supervisory Control and Data Acquisition (SCADA)
system, or Distributed Control System (DCS). You can also use this port to permanently
connect to a maintenance network with Rosemount MON2020.
46
Manual
MS-00809-0100-0470
Emerson.com