Every Two Weeks

Biweekly Cleaning of the Racks and
Figure 15
Crossovers
CROSSOVER
ASSEMBLY (2)
RACK (3)
FEED SHELF
Note
If solution is present in the Bypass, the Silver Recovery Unit is not operating.
[9] Use Kodak SILVER ESTIMATING TEST PAPERS, CAT No. 196 5466, to check the silver content of the
solution flowing into the drain.
[10] If the test shows more than 1 gram/litre of silver, install a new cartridge in the Silver Recovery Unit.
18
DRYER
ASSEMBLY
DETECTOR
CROSSOVER
ASSEMBLY
H112_0066CCA
H112_0066CC
Use only qualified service personnel to perform
all necessary adjustments and procedures.
Warning
Dangerous Voltage. Disconnect the main power before
doing any of the following cleanup procedures.
[1] Perform the Daily Maintenance. See Page 17.
[2] Remove the Dryer Assembly.
Caution
Do not allow water to drip onto the equipment.
[3] Clean any debris or buildup from the dryer area
and the Dryer Assembly with a warm, damp
sponge or cloth.
Caution
• Fixer will contaminate the developer. When
removing the Rack Assemblies, use the Splash
Guard between the developer and fixer Tanks.
• Use the Drip Tray when removing or installing the
Racks. See Figure 9 on Page 9.
• Do not rinse the Detector Crossover Assembly or
the Dryer Assembly.
[4] Remove the Racks.
[5] Rinse the Racks and wipe them with a cloth.
[6] Check the following:
(a) that the Chains move freely and are adjusted
correctly
(b) that all the Rollers rotate correctly
(c) that the space between the Turnaround Side
Plates and the Rack Side Plates is equal and
that the Side Plates are parallel on both ends
[7] Check the operation of the Silver Recovery Unit,
if installed.
[8] Check that no solution is flowing through the
Bypass.