To Test The Pumping System - Agilent Technologies TPS-compact User Manual

Hide thumbs Also See for TPS-compact:
Table of Contents

Advertisement

WARNING!
If dangerous gases were being pumped, flush the pump with air or inert gas
for at least 10 minutes prior to disassembly.

To test the Pumping System

1
2
3
Newly installed tip seals may require several hours of run time to seat properly
NOTE
and enable the pump to meet speed and base pressure specifications.
TPS-compact User Manual / 87-901-010-01 (H)
Run the TPS-compact for about 5 seconds. Verify that the front
fan is running. If you hear loud noises from the Scroll pump this
indicates that the tip seal or main O-ring are possibly out of
place.
Disassemble and repair as necessary.
The pump is now ready to return to service.